Use of a wireless implantable cardioverter defibrillator in a patient with a preexisting neurostimulator
- PMID: 19545350
- DOI: 10.1111/j.1540-8159.2009.02374.x
Use of a wireless implantable cardioverter defibrillator in a patient with a preexisting neurostimulator
Abstract
We report the case of a 54-year-old man with a previously implanted neurostimulator who presented with palpitations and was found to have sustained ventricular tachycardia on electrophysiologic study. A Medtronic wireless implantable cardioverter defibrillator (ICD; Medtronic Inc., Minneapolis, MN, USA) with home monitoring (HM) was successfully implanted. Interaction testing during implantation, follow-up, and HM showed that there was no device-device interaction.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources