Nontraditional surgical approaches for implantation of pacemaker and cardioverter defibrillator systems in patients with limited venous access
- PMID: 19559206
- DOI: 10.1016/j.athoracsur.2009.04.006
Nontraditional surgical approaches for implantation of pacemaker and cardioverter defibrillator systems in patients with limited venous access
Abstract
Background: Indications for placement of implantable cardioverter-defibrillators (ICD) and pacemakers have expanded, and traditional transvenous implantation may not be feasible in patients with aberrant anatomy or venous obstruction. In these settings, successful lead placement has required innovative surgical approaches. A case series of successful placement of these systems in challenging patients is presented.
Methods: A 2-year retrospective study of patients undergoing placement of minimally invasive epicardial pacing leads or ICD coils was performed.
Results: Eleven patients underwent minimally invasive surgical placement of leads or coils. None were converted to open sternotomy. One required extension to minianterior thoracotomy. Causes of intravenous placement failure included aberrant anatomy with failure to access coronary sinus in 9 and venous occlusion in 2. Four patients had previous operations through a median sternotomy. Procedures included left video-assisted thoracoscopic (VATS) placement of a left ventricular epicardial lead in 8, left VATS conversion to minianterior thoracotomy left ventricular epicardial lead placement in 1, left VATS placement of ICD coil in 1, subxiphoid placement of a right ventricular epicardial lead in 1, subxiphoid ICD coil in 2, and subcutaneous ICD coil placement in 3. Mean hospitalization was 4.6 days. Postoperative hypotension and pulmonary edema occurred in 27% of patients. No patients died.
Conclusions: Conventional transvenous lead implantation may be difficult or impossible in some patients with aberrant or occluded venous access. Novel surgical approaches with the use of minimally invasive procedures can establish optimally functional pacing and ICD systems without sternotomy and low associated morbidity.
Similar articles
-
Innovative techniques for placement of implantable cardioverter-defibrillator leads in patients with limited venous access to the heart.Pacing Clin Electrophysiol. 2006 Feb;29(2):181-7. doi: 10.1111/j.1540-8159.2006.00314.x. Pacing Clin Electrophysiol. 2006. PMID: 16492305
-
Nontraditional implantable cardioverter defibrillator placement in adult patients with limited venous access: a case series.Pacing Clin Electrophysiol. 2010 Feb;33(2):217-25. doi: 10.1111/j.1540-8159.2009.02577.x. Epub 2009 Oct 5. Pacing Clin Electrophysiol. 2010. PMID: 19804488
-
Novel minimally invasive, intrapericardial implantable cardioverter defibrillator coil system: a useful approach to arrhythmia therapy in children.Ann Thorac Surg. 2009 Apr;87(4):1234-8; discussion 1238-9. doi: 10.1016/j.athoracsur.2009.01.015. Ann Thorac Surg. 2009. PMID: 19324158
-
Efficacy of prophylactic epicardial pacing leads in children and young adults.Ann Thorac Surg. 2004 Jul;78(1):197-202; discussion 202-3. doi: 10.1016/j.athoracsur.2004.02.008. Ann Thorac Surg. 2004. PMID: 15223427 Review.
-
Completely epicardial implantable cardioverter/defibrillator (ICD) and CRT-D systems: A case series and systematic literature review.Pacing Clin Electrophysiol. 2021 Sep;44(9):1616-1630. doi: 10.1111/pace.14318. Epub 2021 Aug 2. Pacing Clin Electrophysiol. 2021. PMID: 34287989
Cited by
-
Nontraditional placement of an implantable cardioverter-defibrillator in a heterotaxy patient after the completion of total cavopulmonary connection.J Artif Organs. 2013 Dec;16(4):495-7. doi: 10.1007/s10047-013-0714-4. Epub 2013 Jun 1. J Artif Organs. 2013. PMID: 23728474
-
Nothing inside the heart - Combining epicardial pacing with the S-ICD.HeartRhythm Case Rep. 2015 Sep 4;1(6):419-423. doi: 10.1016/j.hrcr.2015.04.001. eCollection 2015 Nov. HeartRhythm Case Rep. 2015. PMID: 28491598 Free PMC article. No abstract available.
-
His-Bundle Pacing in a Patient With Tricuspid and Mitral Prosthetic Valves Without Suitable Coronary Veins for Lead Placement.CJC Open. 2021 Apr 8;3(8):1082-1084. doi: 10.1016/j.cjco.2021.03.012. eCollection 2021 Aug. CJC Open. 2021. PMID: 34505048 Free PMC article.
-
Design and Functionality of a Multilumen Thoracic Access Port for Pericardial Access Under Direct Visualization.J Med Device. 2022 Dec 1;16(4):041005. doi: 10.1115/1.4054999. Epub 2022 Jul 26. J Med Device. 2022. PMID: 36353367 Free PMC article.
-
Long-term follow-up of minimally invasive video-assisted thoracoscopic surgery with epicardial radiofrequency ablation for complex cases of inappropriate sinus tachycardia.HeartRhythm Case Rep. 2015 Sep 8;1(6):477-480. doi: 10.1016/j.hrcr.2015.08.001. eCollection 2015 Nov. HeartRhythm Case Rep. 2015. PMID: 28491610 Free PMC article. No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical