. 2009 Sep;407(1-2):36-42.
doi: 10.1016/j.cca.2009.06.023. Epub 2009 Jun 24.

Quantitative UPLC-MS/MS analysis of underivatised amino acids in body fluids is a reliable tool for the diagnosis and follow-up of patients with inborn errors of metabolism

Affiliations

Quantitative UPLC-MS/MS analysis of underivatised amino acids in body fluids is a reliable tool for the diagnosis and follow-up of patients with inborn errors of metabolism

W A Huub Waterval et al. Clin Chim Acta. 2009 Sep.

Abstract

Background: An electro-spray ionisation ultra-performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) application for the quantitative analysis of amino acids was developed. The suitability for the detection and follow-up of patients suffering from inborn errors of metabolism (IEM) was assessed by extensive cross-validation with ion-exchange liquid chromatography (IEX-LC) with post-column ninhydrin derivatisation, participation in external quality control (ERNDIM) and analysis of samples of patients with confirmed IEM.

Methods: Prior to analysis plasma and urine samples were merely diluted 150-fold in mobile phase. Amino acids were detected in the multiple reaction monitoring mode (MRM) in the ESI-positive mode. The analytical results were compared with IEX-LC. External quality control scheme performance is presented.

Results: Comprehensive analysis of amino acids in plasma and urine was achieved with a run-to-run time of 30min. Validation results were satisfactory and there was a very good correlation between UPLC-MS/MS and IEX-LC. Analytical results obtained in the external quality control scheme were essentially the same as those of the other participants. Patients suffering from IEM were readily identified.

Conclusion: UPLC-MS/MS analysis of amino acids in body fluids is rapid, reliable and suitable for the diagnosis and follow-up patients with IEM.