An in vivo comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by using rotary nickel-titanium files
- PMID: 19567315
- DOI: 10.1016/j.joen.2009.04.025
An in vivo comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by using rotary nickel-titanium files
Abstract
Introduction: The purpose of this study was to compare the accuracy of working length (WL) measurements by using the Root ZX II, Apex NRG XFR, and Mini Apex Locator with rotary nickel-titanium (NiTi) instruments.
Methods: Twenty-eight teeth had their WLs determined with each electronic apex locator (EAL) by using 0.04 taper ProFiles sizes 40-20 in a crown-down technique until WL was reached. Four control teeth had their WL determined by using stainless steel hand files. The files were cemented at WL, and the teeth were extracted. The apical 4 mm of each root was shaved to the apical constriction, exposing the file. Photographs were taken under 15x and 30x magnification and projected at 360x and 720x for evaluation.
Results: The accuracy of the Root ZX II, Apex NRG XFR, and Mini Apex Locator in locating the minor diameter within +/-0.5 mm was 50%, 46.43%, and 39.29%, respectively. There was no statistically significant difference between the EALs in locating the minor diameter. The determination of WL by using hand files in the control teeth was more accurate.
Conclusions: The Root ZX II, Apex NRG XFR, and Mini Apex Locator used with rotary NiTi files were able to locate the apical constriction within +/-0.5 mm only 50% or less of the time.
Similar articles
-
A comparison between two electronic apex locators: an ex vivo investigation.Int Endod J. 2007 May;40(5):362-73. doi: 10.1111/j.1365-2591.2006.01229.x. Epub 2007 Feb 19. Int Endod J. 2007. PMID: 17309741
-
Influence of apical constriction diameter on Root ZX apex locator precision.J Endod. 2007 Aug;33(8):995-8. doi: 10.1016/j.joen.2007.05.009. J Endod. 2007. PMID: 17878091
-
The accuracy of the Root ZX electronic apex locator using stainless-steel and nickel-titanium files.J Endod. 2003 Oct;29(10):662-3. doi: 10.1097/00004770-200310000-00012. J Endod. 2003. PMID: 14606791 No abstract available.
-
Electronic apex locator.Dent Clin North Am. 2004 Jan;48(1):35-54. doi: 10.1016/j.cden.2003.10.005. Dent Clin North Am. 2004. PMID: 15066506 Review.
-
Nickel-titanium rotary systems: review of existing instruments and geometries.Dent Today. 2000 Oct;19(10):86-8, 90-5. Dent Today. 2000. PMID: 12524811 Review. No abstract available.
Cited by
-
Comparison of the cyclic fatigue resistance of One Curve, F6 Skytaper, Protaper Next, and Hyflex CM endodontic files.Restor Dent Endod. 2022 Mar 4;47(2):e16. doi: 10.5395/rde.2022.47.e16. eCollection 2022 May. Restor Dent Endod. 2022. PMID: 35692224 Free PMC article.
-
Foramen Changes following Over Instrumentation of Curved Canals with Three Engine-Driven Instruments: An In Vitro Study.Iran Endod J. 2017 Fall;12(4):454-461. doi: 10.22037/iej.v12i4.16268. Iran Endod J. 2017. PMID: 29225641 Free PMC article.
-
Evaluation of Two New Electronic Apex-Locator-Controlled Handpieces Using a NiTi Rotary File: An In Vitro Study.J Dent (Tehran). 2013 Nov;10(6):501-5. Epub 2013 Nov 30. J Dent (Tehran). 2013. PMID: 24910660 Free PMC article.
-
An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files.Clin Cosmet Investig Dent. 2018 May 22;10:75-82. doi: 10.2147/CCIDE.S158882. eCollection 2018. Clin Cosmet Investig Dent. 2018. PMID: 29861643 Free PMC article.
-
An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files.Restor Dent Endod. 2016 Feb;41(1):6-11. doi: 10.5395/rde.2016.41.1.6. Epub 2016 Jan 4. Restor Dent Endod. 2016. PMID: 26877985 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous