Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2009 Jun;17(6):1281-6.
doi: 10.1038/oby.2009.14. Epub 2009 Feb 19.

Body composition measured by dual-energy X-ray absorptiometry half-body scans in obese adults

Affiliations
Comparative Study

Body composition measured by dual-energy X-ray absorptiometry half-body scans in obese adults

Megan P Rothney et al. Obesity (Silver Spring). 2009 Jun.

Abstract

Dual-energy X-ray absorptiometry (DXA) has become a common measurement of human body composition. However, obese subjects have been understudied largely due to weight and scan area restrictions. Newer DXA instruments allow for heavier subjects to be supported by the DXA scanner, but the imaging area is still smaller than the body size of some obese subjects. In this study, we determined the validity of an automated half-scan methodology by comparing to the standard whole-body scans in a cohort of obese volunteers. Fifty-two subjects whose BMI>30 kg/m2 completed whole-body iDXA (GE Lunar) scans. The resulting scans were analyzed in three ways: the standard whole-body scan, total body estimated from the left side, and from the right side. Fat mass, nonbone lean mass, bone mineral content (BMC), and percent fat derived from each half scan were compared to the whole-body scans. Total fat mass, nonbone lean mass, or percent fat was comparable for the whole-body scans, left, and right side scans (>97% within individuals and >99.9% for the group). The BMC estimate using the right side scan was slightly but statistically higher than the whole-body BMC (approximately 30 g or 1%, P<0.001), while the left side scan BMC estimate was lower than the whole-body BMC by the same magnitude. No significant magnitude bias was found for any of the composition variables. We conclude that the new iDXA half-body analysis in obese subjects appears to be closely comparable to whole-body analysis for fat mass, nonbone lean mass, and percent fat.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Example of iDXA total-body composition image and a representation of the portions of the total-body image used to generate the right side and left side images.
Figure 2
Figure 2
Correlation plots between right side, left side, and whole-body scans (n = 52). Dashed lines represent line of identity.
Figure 3
Figure 3
Bland–Altman plots comparing the right side, left side, and whole-body scans. No significant magnitude bias was detected for any of the composition parameters. All r2 < 0.033 (not significant, n = 52).

References

    1. Lee S, Gallagher D. Assessment methods in human body composition. Curr Opin Clin Nutr Metab Care. 2008;11:566–572. - PMC - PubMed
    1. Ellis KJ. Human body composition: in vivo methods. Physiol Rev. 2000;80:649–680. - PubMed
    1. Goodsitt MM. Evaluation of a new set of calibration standards for the measurement of fat content via DPA and DXA. Med Phys. 1992;19:35–44. - PubMed
    1. Tataranni PA, Ravussin E. Use of dual-energy X-ray absorptiometry in obese individuals. Am J Clin Nutr. 1995;62:730–734. - PubMed
    1. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310. - PubMed

Publication types