Comparative Study
. 2009 Sep;47(9):2699-703.
doi: 10.1128/JCM.00762-09. Epub 2009 Jul 8.

Field performance of a rapid diagnostic test for influenza in an ambulatory setting

Affiliations
Comparative Study

Field performance of a rapid diagnostic test for influenza in an ambulatory setting

Isabelle Rouleau et al. J Clin Microbiol. 2009 Sep.

Abstract

Provided test characteristics are adequate, point-of-care rapid antigen detection tests for influenza could improve the timeliness and appropriateness of clinical decisions. Our objective was to estimate the field sensitivity and specificity of the Quidel QuickVue Influenza A+B test in an ambulatory setting. The sensitivity and specificity of the Quidel QuickVue test was evaluated against reverse-transcriptase PCR (RT-PCR) on nasopharyngeal specimens collected over two consecutive influenza seasons from ambulatory patients consulting for influenza-like illness (ILI) within 7 days of ILI onset. A total of 491 patients with ILI (180 in 2006 to 2007 and 311 in 2007 to 2008) provided specimens that were tested both by PCR and by the Quidel QuickVue test. Among the 267 patients positive by PCR (55%), 52 were also positive by the QuickVue test, for an overall sensitivity of 19.5% (95% confidence interval [95% CI], 14.7% to 24.2%). Among the 221 PCR-negative patients, 2 were positive for influenza B virus by the rapid test (<1%), for an overall specificity of 99.1% (95% CI, 97.9 to 100%). The field sensitivity of the test varied little with the age or gender of the patient, immunization status, delay since the onset of symptoms, or influenza season. The sensitivity of the test was slightly but nonsignificantly higher for influenza B virus (23%) than for influenza A virus (18%). Despite its high specificity, the low sensitivity of the Quidel QuickVue Influenza A+B test is too poor to direct clinical decisions for ambulatory patients with ILI. Negative results cannot rule out the diagnosis of influenza, and in that context, this test is of questionable utility for routine application in the clinical setting.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Barenfanger, J., C. Drake, N. Leon, T. Mueller, and T. Troutt. 2000. Clinical and financial benefits of rapid detection of respiratory viruses: an outcomes study. J. Clin. Microbiol. 382824-2828. - PMC - PubMed
    1. Bellei, N., D. Benfica, A. H. Perosa, R. Carlucci, M. Barros, and C. Granato. 2003. Evaluation of a rapid test (QuickVue) compared with the shell vial assay for detection of influenza virus clearance after antiviral treatment. J. Virol. Methods 10985-88. - PubMed
    1. Byington, C. L., H. Castillo, K. Gerber, J. A. Daly, L. A. Brimley, S. Adams, J. C. Christenson, and A. T. Pavia. 2002. The effect of rapid respiratory viral diagnostic testing on antibiotic use in a children's hospital. Arch. Pediatr. Adolesc. Med. 1561230-1234. - PubMed
    1. Cazacu, A. C., J. Greer, M. Taherivand, and G. J. Demmler. 2003. Comparison of lateral-flow immunoassay and enzyme immunoassay with viral culture for rapid detection of influenza virus in nasal wash specimens from children. J. Clin. Microbiol. 412132-2134. - PMC - PubMed
    1. Covalciuc, K. A., K. H. Webb, and C. A. Carlson. 1999. Comparison of four clinical specimen types for detection of influenza A and B viruses by optical immunoassay (FLU OIA test) and cell culture methods. J. Clin. Microbiol. 373971-3974. - PMC - PubMed

MeSH terms

Substances