Mass cytometry: technique for real time single cell multitarget immunoassay based on inductively coupled plasma time-of-flight mass spectrometry
- PMID: 19601617
- DOI: 10.1021/ac901049w
Mass cytometry: technique for real time single cell multitarget immunoassay based on inductively coupled plasma time-of-flight mass spectrometry
Abstract
A novel instrument for real time analysis of individual biological cells or other microparticles is described. The instrument is based on inductively coupled plasma time-of-flight mass spectrometry and comprises a three-aperture plasma-vacuum interface, a dc quadrupole turning optics for decoupling ions from neutral components, an rf quadrupole ion guide discriminating against low-mass dominant plasma ions, a point-to-parallel focusing dc quadrupole doublet, an orthogonal acceleration reflectron analyzer, a discrete dynode fast ion detector, and an 8-bit 1 GHz digitizer. A high spectrum generation frequency of 76.8 kHz provides capability for collecting multiple spectra from each particle-induced transient ion cloud, typically of 200-300 micros duration. It is shown that the transients can be resolved and characterized individually at a peak frequency of 1100 particles per second. Design considerations and optimization data are presented. The figures of merit of the instrument are measured under standard inductively coupled plasma (ICP) operating conditions (<3% cerium oxide ratio). At mass resolution (full width at half-maximum) M/DeltaM > 900 for m/z = 159, the sensitivity with a standard sample introduction system of >1.4 x 10(8) ion counts per second per mg L(-1) of Tb and an abundance sensitivity of (6 x 10(-4))-(1.4 x 10(-3)) (trailing and leading masses, respectively) are shown. The mass range (m/z = 125-215) and abundance sensitivity are sufficient for elemental immunoassay with up to 60 distinct available elemental tags. When <15 elemental tags are used, a higher sensitivity mode at lower resolution (M/DeltaM > 500) can be used, which provides >2.4 x 10(8) cps per mg L(-1) of Tb, at (1.5 x 10(-3))-(5.0 x 10(-3)) abundance sensitivity. The real-time simultaneous detection of multiple isotopes from individual 1.8 microm polystyrene beads labeled with lanthanides is shown. A real time single cell 20 antigen expression assay of model cell lines and leukemia patient samples immuno-labeled with lanthanide-tagged antibodies is presented.
Similar articles
-
Sensitive sandwich immunoassay based on single particle mode inductively coupled plasma mass spectrometry detection.Talanta. 2010 Nov 15;83(1):48-54. doi: 10.1016/j.talanta.2010.08.037. Epub 2010 Sep 27. Talanta. 2010. PMID: 21035642
-
Protein phosphorylation stoichiometry by simultaneous ICP-QMS determination of phosphorus and sulfur oxide ions: a multivariate optimization of plasma operating conditions.Talanta. 2010 Feb 15;80(4):1513-25. doi: 10.1016/j.talanta.2009.06.082. Epub 2009 Jul 10. Talanta. 2010. PMID: 20082809
-
Isotope ratio determination in boron analysis.Biol Trace Elem Res. 1998 Winter;66(1-3):39-53. doi: 10.1007/BF02783124. Biol Trace Elem Res. 1998. PMID: 10050906 Review.
-
Determination of ion and neutral loss compositions and deconvolution of product ion mass spectra using an orthogonal acceleration time-of-flight mass spectrometer and an ion correlation program.Rapid Commun Mass Spectrom. 2006;20(2):89-102. doi: 10.1002/rcm.2277. Rapid Commun Mass Spectrom. 2006. PMID: 16331746
-
Inductively coupled plasma mass spectrometry-based immunoassay: a review.Mass Spectrom Rev. 2014 Sep-Oct;33(5):373-93. doi: 10.1002/mas.21391. Epub 2013 Nov 22. Mass Spectrom Rev. 2014. PMID: 24272753 Review.
Cited by
-
Comprehensive single-shot biophysical cytometry using simultaneous quantitative phase imaging and Brillouin spectroscopy.Sci Rep. 2022 Oct 31;12(1):18285. doi: 10.1038/s41598-022-23049-4. Sci Rep. 2022. PMID: 36316372 Free PMC article.
-
A 34-Marker Panel for Imaging Mass Cytometric Analysis of Human Snap-Frozen Tissue.Front Immunol. 2020 Jul 16;11:1466. doi: 10.3389/fimmu.2020.01466. eCollection 2020. Front Immunol. 2020. PMID: 32765508 Free PMC article.
-
Systematic Study of Immune Cell Diversity in ischemic postconditioning Using High-Dimensional Single-Cell Analysis with Mass Cytometry.Aging Dis. 2021 Jun 1;12(3):812-825. doi: 10.14336/AD.2020.1115. eCollection 2021 Jun. Aging Dis. 2021. PMID: 34094644 Free PMC article.
-
New assays for monitoring residual HIV burden in effectively treated individuals.Curr Opin HIV AIDS. 2013 Mar;8(2):106-10. doi: 10.1097/COH.0b013e32835d811b. Curr Opin HIV AIDS. 2013. PMID: 23314907 Free PMC article. Review.
-
Cytofkit: A Bioconductor Package for an Integrated Mass Cytometry Data Analysis Pipeline.PLoS Comput Biol. 2016 Sep 23;12(9):e1005112. doi: 10.1371/journal.pcbi.1005112. eCollection 2016 Sep. PLoS Comput Biol. 2016. PMID: 27662185 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical