Victim of carbon monoxide poisoning identified by carbon monoxide oximetry
- PMID: 19615844
- DOI: 10.1016/j.jemermed.2009.05.017
Victim of carbon monoxide poisoning identified by carbon monoxide oximetry
Abstract
Background: Carbon monoxide (CO) is known as a leading cause of unintentional poisoning death in many countries. Diagnosis is usually made by measuring carboxyhemoglobin (COHb), but due to the non-specific symptoms, the crucial step is considering CO poisoning. A possible solution might be screening emergency department (ED) patients. A cutaneous sensor that measures oxygen and CO saturation simultaneously could accomplish that.
Objectives: We report a case in which CO-oximetry screening helped to identify CO poisoning in a patient with otherwise non-specific complaints presenting to a busy university hospital ED.
Case report: A 53-year-old woman presented to our ED, reporting non-specific but common symptoms including emesis and diarrhea, one-sided headache, paresthesia, and palpitations. Whereas conventional pulse oximetry showed normal results, a COHb level of 28% was measured using a CO-oximeter. This allowed a diagnosis of CO poisoning to be made at admission. The patient was treated with 100% oxygen continuous positive airway pressure ventilation, and CO washout was monitored continuously and non-invasively. In the meantime, the local fire department was alerted and found a malfunctioning water heater as the source of CO in the patient's residence. As COHb fell to normal levels and the patient reported feeling well, she was discharged the next day.
Conclusion: CO-oximetry served as a powerful tool in both identifying and monitoring a CO-poisoned patient in a busy ED. Future studies should address possible sources of error using this method on a long-term basis and its level of acceptance by ED staff.
Copyright © 2011 Elsevier Inc. All rights reserved.
Similar articles
-
Non-invasive pulse CO-oximetry screening in the emergency department identifies occult carbon monoxide toxicity.J Emerg Med. 2008 May;34(4):441-50. doi: 10.1016/j.jemermed.2007.12.004. Epub 2008 Jan 28. J Emerg Med. 2008. PMID: 18226877
-
Finding needles in a haystack: a case series of carbon monoxide poisoning detected using new technology in the emergency department.Clin Toxicol (Phila). 2008 Jun;46(5):461-9. doi: 10.1080/15563650701725110. Clin Toxicol (Phila). 2008. PMID: 18568803
-
Emergency department management of suspected carbon monoxide poisoning: role of pulse CO-oximetry.Respir Care. 2013 Oct;58(10):1614-20. doi: 10.4187/respcare.02313. Epub 2013 Mar 19. Respir Care. 2013. PMID: 23513247
-
Carbon monoxide poisoning.Crit Care Clin. 2012 Oct;28(4):537-48. doi: 10.1016/j.ccc.2012.07.007. Crit Care Clin. 2012. PMID: 22998990 Review.
-
Diagnosis and management of carbon monoxide poisoning in the emergency department.Emerg Med Pract. 2011 Feb;13(2):1-14; quiz 14. Emerg Med Pract. 2011. PMID: 22164402 Review.
Cited by
-
Carboxyhemoglobinemia in Critically Ill Coronavirus Disease 2019 Patients.J Clin Med. 2021 Jun 21;10(12):2731. doi: 10.3390/jcm10122731. J Clin Med. 2021. PMID: 34205655 Free PMC article.
-
Assessment of absorbed dose of gamma rays using the simultaneous determination of inactive hemoglobin derivatives as a biological dosimeter.Radiat Environ Biophys. 2020 Mar;59(1):131-144. doi: 10.1007/s00411-019-00821-1. Epub 2019 Nov 16. Radiat Environ Biophys. 2020. PMID: 31734721
-
Can indicators of myocardial damage predict carbon monoxide poisoning outcomes?BMC Emerg Med. 2021 Jan 15;21(1):7. doi: 10.1186/s12873-021-00405-7. BMC Emerg Med. 2021. PMID: 33451295 Free PMC article.
-
Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.Anesth Analg. 2013 Oct;117(4):847-858. doi: 10.1213/ANE.0b013e31828610a0. Epub 2013 Mar 11. Anesth Analg. 2013. PMID: 23477959 Free PMC article. Clinical Trial.
-
Diagnosis of carbon monoxide exposure in clinical research and practice: A scoping review.PLoS One. 2025 Feb 5;20(2):e0300989. doi: 10.1371/journal.pone.0300989. eCollection 2025. PLoS One. 2025. PMID: 39908298 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials