The impact of reconstruction method on the quantification of DaTSCAN images
- PMID: 19618181
- DOI: 10.1007/s00259-009-1212-z
The impact of reconstruction method on the quantification of DaTSCAN images
Abstract
Purpose: Reconstruction of DaTSCAN brain studies using OS-EM iterative reconstruction offers better image quality and more accurate quantification than filtered back-projection. However, reconstruction must proceed for a sufficient number of iterations to achieve stable and accurate data. This study assessed the impact of the number of iterations on the image quantification, comparing the results of the iterative reconstruction with filtered back-projection data.
Methods: A striatal phantom filled with (123)I using striatal to background ratios between 2:1 and 10:1 was imaged on five different gamma camera systems. Data from each system were reconstructed using OS-EM (which included depth-independent resolution recovery) with various combinations of iterations and subsets to achieve up to 200 EM-equivalent iterations and with filtered back-projection. Using volume of interest analysis, the relationships between image reconstruction strategy and quantification of striatal uptake were assessed.
Results: For phantom filling ratios of 5:1 or less, significant convergence of measured ratios occurred close to 100 EM-equivalent iterations, whereas for higher filling ratios, measured uptake ratios did not display a convergence pattern. Assessment of the count concentrations used to derive the measured uptake ratio showed that nonconvergence of low background count concentrations caused peaking in higher measured uptake ratios. Compared to filtered back-projection, OS-EM displayed larger uptake ratios because of the resolution recovery applied in the iterative algorithm.
Conclusion: The number of EM-equivalent iterations used in OS-EM reconstruction influences the quantification of DaTSCAN studies because of incomplete convergence and possible bias in areas of low activity due to the nonnegativity constraint in OS-EM reconstruction. Nevertheless, OS-EM using 100 EM-equivalent iterations provides the best linear discriminatory measure to quantify the uptake in DaTSCAN studies.
Similar articles
-
Optimisation of the OS-EM algorithm and comparison with FBP for image reconstruction on a dual-head camera: a phantom and a clinical 18F-FDG study.Eur J Nucl Med Mol Imaging. 2003 Nov;30(11):1510-9. doi: 10.1007/s00259-003-1246-6. Epub 2003 Sep 23. Eur J Nucl Med Mol Imaging. 2003. PMID: 14579091 Clinical Trial.
-
Iterative three-dimensional expectation maximization restoration of single photon emission computed tomography images: application in striatal imaging.Med Phys. 2006 Jan;33(1):52-60. doi: 10.1118/1.2135908. Med Phys. 2006. PMID: 16485409
-
Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system.J Nucl Med. 2005 May;46(5):868-77. J Nucl Med. 2005. PMID: 15872362 Clinical Trial.
-
Characterization of ordered-subsets expectation maximization with 3D post-reconstruction Gauss filtering and comparison with filtered backprojection in 99mTc SPECT.Ann Nucl Med. 2005 Apr;19(2):75-82. doi: 10.1007/BF03027384. Ann Nucl Med. 2005. PMID: 15909485
-
Absolute quantification in dopaminergic neurotransmission SPECT using a Monte Carlo-based scatter correction and fully 3-dimensional reconstruction.J Nucl Med. 2005 Sep;46(9):1497-504. J Nucl Med. 2005. PMID: 16157533
Cited by
-
Design of a Multi-Pinhole Collimator for I-123 DaTscan Imaging on Dual-Headed SPECT Systems in Combination with a Fan-Beam Collimator.IEEE Trans Nucl Sci. 2016 Feb;63(1):90-97. doi: 10.1109/TNS.2016.2515519. Epub 2016 Feb 3. IEEE Trans Nucl Sci. 2016. PMID: 27182078 Free PMC article.
-
Comparison of machine learning and semi-quantification algorithms for (I123)FP-CIT classification: the beginning of the end for semi-quantification?EJNMMI Phys. 2017 Nov 29;4(1):29. doi: 10.1186/s40658-017-0196-1. EJNMMI Phys. 2017. PMID: 29188397 Free PMC article.
-
Development of a novel phantom for evaluating jawbone SPECT targeting medication-related osteonecrosis of the jaw.Oral Radiol. 2025 Jul;41(3):378-387. doi: 10.1007/s11282-025-00809-2. Epub 2025 Feb 27. Oral Radiol. 2025. PMID: 40014280 Free PMC article.
-
A new visual rating scale for Ioflupane imaging in Lewy body disease.Neuroimage Clin. 2018;20:823-829. doi: 10.1016/j.nicl.2018.09.012. Epub 2018 Sep 20. Neuroimage Clin. 2018. PMID: 30268991 Free PMC article.
-
Robust, fully automatic delineation of the head contour by stereotactical normalization for attenuation correction according to Chang in dopamine transporter scintigraphy.Eur Radiol. 2015 Sep;25(9):2709-17. doi: 10.1007/s00330-015-3667-6. Epub 2015 Feb 26. Eur Radiol. 2015. PMID: 25716941
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources