Regional anaesthesia elastomeric pump performance after a single use and subsequent refill: a laboratory study
- PMID: 19624633
- DOI: 10.1111/j.1365-2044.2009.05941.x
Regional anaesthesia elastomeric pump performance after a single use and subsequent refill: a laboratory study
Abstract
Ambulatory local anaesthetic delivery systems are often limited by a short effective duration of infusion. Prolonging nerve blockade by substituting a new pump as recommended by the manufacturers, represents a substantial consumable item cost ($US 300-500). We therefore evaluated the flow delivery performance of 31 single model elastomeric devices (all with a 2 ml.h(-1) background and 5 ml every hour bolus capability) that had been filled, used in clinical practice and then refilled in the laboratory. For the second infusion, there was a pattern of over-infusion (< 10 ml.h(-1)) in the first hour; however, all pumps depreciated to < 150% of predicted by the second hour. The subsequent performance of all pumps was not only within safe limits, but also predominantly within the range (background infusion +/- 15%, bolus +10/-20%) specified by the manufacturer for primary infusion. We conclude that this elastomeric regional anaesthesia pump design performs satisfactorily after having been refilled following a single previous use.
Similar articles
-
Elastomeric pump reliability in postoperative regional anesthesia: a survey of 430 consecutive devices.Anesth Analg. 2008 Dec;107(6):2079-84. doi: 10.1213/ane.0b013e318187c9bb. Anesth Analg. 2008. PMID: 19020162
-
Continuous postoperative pain control after bone graft harvest using a pump delivery system.J Oral Maxillofac Surg. 2001 Jun;59(6):707-8. doi: 10.1053/joms.2001.23414. J Oral Maxillofac Surg. 2001. PMID: 11381403 No abstract available.
-
A comparison of the hourly output between the Ambu® Smart-Infuser™ Pain Pump and the On-Q Pump® with Select-A-Flow™ Variable Rate Controller with standard and overfill volumes.Paediatr Anaesth. 2016 Apr;26(4):425-8. doi: 10.1111/pan.12850. Epub 2016 Jan 20. Paediatr Anaesth. 2016. PMID: 26786058
-
High-technology i.v. infusion devices.Am J Hosp Pharm. 1989 Feb;46(2):320-35. Am J Hosp Pharm. 1989. PMID: 2653027 Review.
-
Future of regional anaesthesia.Acta Anaesthesiol Scand. 2005 Aug;49(7):913-8. doi: 10.1111/j.1399-6576.2005.00774.x. Acta Anaesthesiol Scand. 2005. PMID: 16045649 Review.