Material properties of a maxillofacial chlorinated polyethylene elastomer stored in simulated skin secretions

Affiliations

Material properties of a maxillofacial chlorinated polyethylene elastomer stored in simulated skin secretions

Panagiota N Eleni et al. J Biomed Mater Res B Appl Biomater. 2009 Nov.

Abstract

Statement of problem: Facial prostheses deteriorate in a service environment primary due to exposition to various environmental factors, including sebaceous oils (sebum) and perspiration.

Purpose: This study investigated the physical properties of an experimental, facial prosthetic after immersion for 6 months in simulated sebum, and perspiration at 37 degrees C.

Material and methods: Chlorinated polyethylene (CPE) specimens were immersed in simulated perspiration as well as in sebum. Compression tests were conducted on a Zwick testing machine. Shore A hardness measurements were carried out in a CV digital Shore A durometer. Melting and glass transition temperatures were evaluated with a differential scanning calorimeter. Weight changes were measured and color changes were determined in the CIE LAB system using a MiniScan XE spectrophotometer. Simple mathematical models were developed to correlate the measured properties with immersion time. The data were also subjected to analyses of variance (ANOVA) and the Tukey multiple range tests at a level of alpha = 0.05.

Results: Specimens immersed in perspiration became harder due to facilitation of the propagation of cross-linking reaction that probably occurred during aging of the CPE samples. Some weight increase was observed for the specimens immersed into the aqueous solutions, whereas for those immersed in sebum, weight loss was recorded, probably because of extraction of some compounds. The color change was higher for the specimens immersed in sebum than that corresponding to simulated perspiration.

Conclusions: The chlorinated polyethylene specimens aged for a period, which simulates 1.5 years of clinical service1, showed significant deformations in their physical properties.

PubMed Disclaimer

References

REFERENCES

    1. Polyzois GL,Tarantili PA,Frangou MJ,Andreopoulos AG. Physical properties of a silicone prosthetic elastomer stored in simulated skin secretions. J Prosthet Dent 2000; 83: 572-577.
    1. Chalian VA,Phillips RW. Materials in maxillofacial prosthetics. J Biomed Mater Res Symp 1974; 8: 349-363.
    1. Abdelnnabi MM,Moore DJ,Sakumura JS. In vitro comparison study of MDX-4-4210 and polydimethyl siloxane silicone materials. J Prosthet Dent 1984; 51: 523-526.
    1. Khan Z,Gettleman L,Jacobsen C. Conference report: Materials research in maxillofacial prosthetics. J Dent Res 1992; 71: 1541-1542.
    1. Bellamy KE,Waters MGJ. Designing a prosthesis to simulate the elastic properties of skin. Biomed Mater Eng 2005; 15: 21-27.

LinkOut - more resources