The determination of glutamine with flow-injection chemiluminescence detection and mechanism study
- PMID: 19630088
- DOI: 10.1002/bio.1143
The determination of glutamine with flow-injection chemiluminescence detection and mechanism study
Abstract
The main purpose of this study was to develop an inexpensive, simple, rapid and sensitive chemiluminescence (CL) method for the determination of glutamine (Gln) using a flow-injection (FI) system. Gln was found to strongly inhibit the CL signal of the luminol-H(2)O(2)-CuSO(4) system in Na(2)B(4)O(7) solution. A new FI-CL method was developed for the determination of Gln. Parameters affecting the reproducibility and CL detection were optimized systematically. Under the optimized conditions, the corresponding linear regression equation was established over the range of 5.0 x 10(-7) to 2.5 x 10(-6) mol/L with the detection limit of 1.8 x 10(-8) mol/L. The relative standard deviation was found to be 1.8% for 11 replicate determinations of 1.5 x 10(-6) mol/L Gln. The proposed method has been satisfactorily applied for the determination of Gln in real samples (Marzulene-s granules) with recoveries in the range of 98.7-108.6%. The minimum sampling rate was about 100 samples/h. The possible mechanism of this inhibitory CL was studied by fluorescence spectrophotometer and UV-vis spectrophotometer.
Copyright (c) 2009 John Wiley & Sons, Ltd.
Similar articles
-
Sensitive determination of epinephrine in pharmaceutical preparation by flow injection coupled with chemiluminescence detection and mechanism study.Luminescence. 2011 Jan-Feb;26(1):59-64. doi: 10.1002/bio.1186. Epub 2009 Dec 16. Luminescence. 2011. PMID: 20017128
-
Determination of bisphenol A using a flow injection inhibitory chemiluminescence method.Luminescence. 2005 Jan-Feb;20(1):46-50. doi: 10.1002/bio.804. Luminescence. 2005. PMID: 15685659
-
Determination of trace amounts of dopamine by flow-injection analysis coupled with luminol-Ag(III) complex chemiluminescence detection.Luminescence. 2011 Mar-Apr;26(2):93-100. doi: 10.1002/bio.1189. Luminescence. 2011. PMID: 20446341
-
Determination of tetracycline, chlortetracycline and oxytetracycline by flow injection with inhibitory chemiluminescence detection using copper(II) as a probe ion.Luminescence. 2006 Mar-Apr;21(2):106-11. doi: 10.1002/bio.893. Luminescence. 2006. PMID: 16416504
-
A sensitive inhibition chemiluminescence method for the determination of trace tannic acid using the reaction of luminol-hydrogen peroxide catalysed by tetrasulphonated manganese phthalocyanine.Luminescence. 2007 Jan-Feb;22(1):46-52. doi: 10.1002/bio.925. Luminescence. 2007. PMID: 16874849
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources