Vessel calibre and haemoglobin effects on pulse oximetry
- PMID: 19636087
- DOI: 10.1088/0967-3334/30/9/001
Vessel calibre and haemoglobin effects on pulse oximetry
Erratum in
- Physiol Meas. 2010 May;31(5):727
Abstract
Despite its success as a clinical monitoring tool, pulse oximetry may be improved with respect to the need for empirical calibration and the reports of biases in readings associated with peripheral vasoconstriction and haemoglobin concentration. To effect this improvement, this work aims to improve the understanding of the photoplethysmography signal-as used by pulse oximeters-and investigates the effect of vessel calibre and haemoglobin concentration on pulse oximetry. The digital temperature and the transmission of a wide spectrum of light through the fingers of 57 people with known haemoglobin concentrations were measured and simulations of the transmission of that spectrum of light through finger models were performed. Ratios of pulsatile attenuations of light as used in pulse oximetry were dependent upon peripheral temperature and on blood haemoglobin concentration. In addition, both the simulation and in vivo results showed that the pulsatile attenuation of light through fingers was approximately proportional to the absorption coefficients of blood, only when the absorption coefficients were small. These findings were explained in terms of discrete blood vessels acting as barriers to light transmission through tissue. Due to the influence of discrete blood vessels on light transmission, pulse oximeter outputs tend to be dependent upon haemoglobin concentration and on the calibre of pulsing blood vessels-which are affected by vasoconstriction/vasodilation. The effects of discrete blood vessels may account for part of the difference between the Beer-Lambert pulse oximetry model and empirical calibration.
Similar articles
-
Pulse oximetry in the oesophagus.Physiol Meas. 2006 Jan;27(1):R1-35. doi: 10.1088/0967-3334/27/1/R01. Epub 2005 Nov 29. Physiol Meas. 2006. PMID: 16365505 Review.
-
Simple photon diffusion analysis of the effects of multiple scattering on pulse oximetry.IEEE Trans Biomed Eng. 1991 Dec;38(12):1194-203. doi: 10.1109/10.137285. IEEE Trans Biomed Eng. 1991. PMID: 1774081
-
Non-invasive measurement of peripheral venous oxygen saturation using a new venous oximetry method: evaluation during bypass in heart surgery.Physiol Meas. 2007 Aug;28(8):897-911. doi: 10.1088/0967-3334/28/8/012. Epub 2007 Jul 19. Physiol Meas. 2007. PMID: 17664681
-
Changes in the arteriolar volume pulse of the finger during various degrees of tilt using near infra-red and red photoplethysmography.Eur J Med Res. 1998 May 12;3(5):249-55. Eur J Med Res. 1998. PMID: 9580571
-
Pulse oximetry: evolution and directions.Int J Clin Pract. 2000 Sep;54(7):438-44. Int J Clin Pract. 2000. PMID: 11070568 Review.
Cited by
-
Noninvasive hemoglobin sensing and imaging: optical tools for disease diagnosis.J Biomed Opt. 2022 Aug;27(8):080901. doi: 10.1117/1.JBO.27.8.080901. J Biomed Opt. 2022. PMID: 35922891 Free PMC article.
-
Novel wearable and wireless ring-type pulse oximeter with multi-detectors.Sensors (Basel). 2014 Sep 19;14(9):17586-99. doi: 10.3390/s140917586. Sensors (Basel). 2014. PMID: 25244586 Free PMC article.
-
Prediction of arterial blood gas values from venous blood gas values in Asiatic black bears (Ursus thibetanus) anesthetized with intramuscular medetomidine and zolazepam-tiletamine.J Vet Med Sci. 2017 Oct 20;79(10):1757-1763. doi: 10.1292/jvms.16-0596. Epub 2017 Sep 10. J Vet Med Sci. 2017. PMID: 28890467 Free PMC article.
-
Multispectral tissue characterization for intestinal anastomosis optimization.J Biomed Opt. 2015 Oct;20(10):106001. doi: 10.1117/1.JBO.20.10.106001. J Biomed Opt. 2015. PMID: 26440616 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources