. 2009 Sep;46(Pt 5):407-12.
doi: 10.1258/acb.2009.009038. Epub 2009 Jul 29.

Serum free light chain assay reduces the need for serum and urine immunofixation electrophoresis in the evaluation of monoclonal gammopathy

Affiliations

Serum free light chain assay reduces the need for serum and urine immunofixation electrophoresis in the evaluation of monoclonal gammopathy

Richard B Fulton et al. Ann Clin Biochem. 2009 Sep.

Abstract

Background: The potential for serum free light chain (sFLC) assay measurements to replace urine electrophoresis (uEPG) and to also diminish the need for serum immunofixation (sIFE) in the screening for monoclonal gammopathy was assessed. A testing algorithm for monoclonal protein was developed based on our data and cost analysis.

Methods: Data from 890 consecutive sFLC requests were retrospectively analysed. These included 549 samples for serum electrophoresis (sEPG), 447 for sIFE, and 318 for uEPG and urine immunofixation (uIFE). A total of 219 samples had sFLC, sEPG, sIFE and uEPG + uIFE performed. The ability of different test combinations to detect the presence of monoclonal proteins was compared.

Results: The sFLC kappa/lambda ratio (FLC ratio) indicated monoclonal light chains in 12% more samples than uEPG + uIFE. The combination of sEPG and FLC ratio detected monoclonal proteins in 49% more samples than the combination of sEPG and sIFE. Furthermore, the sEPG + FLC ratio combination detected monoclonal protein in 6% more samples than were detected by the combined performance of sEPG, sIFE, uEPG and uIFE. However, non-linearity of the assay, the expense of repeat determinations due to the narrow measuring ranges, and frequent antigen excess checks were found to be limitations of the sFLC assay in this study.

Conclusion: The FLC ratio is a more sensitive method than uIFE in the detection of monoclonal light chains and may substantially reduce the need for onerous 24 h urine collections. Our proposed algorithm for the evaluation of monoclonal gammopathy incorporates the sFLC assay, resulting in a reduction in the performance of labour intensive sIFE and uEPG + uIFE while still increasing the detection of monoclonal proteins.