Force feedback requirements for efficient laparoscopic grasp control
- PMID: 19662556
- DOI: 10.1080/00140130902912803
Force feedback requirements for efficient laparoscopic grasp control
Abstract
During laparoscopic grasping, tissue damage may occur due to use of excessive grasp forces and tissue slippage, whereas in barehanded grasping, humans control their grasp to prevent slippage and use of excessive force (safe grasp). This study investigates the differences in grasp control during barehanded and laparoscopic lifts. Ten novices performed lifts in order to compare pinch forces under four conditions: barehanded; using tweezers; a low-efficient grasper; and a high-efficient grasper. Results showed that participants increased their pinch force significantly later during a barehanded lift (at a pull-force level of 2.63 N) than when lifting laparoscopically (from pull-force levels of 0.77 to 1.08 N). In barehanded lifts all participants could accomplish a safe grasp, whereas in laparoscopic lifts excessive force (up to 7.9 N) and slippage (up to 38% of the trials) occurred frequently. For novices, it can be concluded that force feedback (additional to the hand-tool interface), as in skin-tissue contact, is a prerequisite to maintain a safe grasp. Much is known about grasp control during barehanded object manipulation, especially the control of pinch forces to changing loading, whereas little is known about force perception and grasp control during tool usage. This knowledge is a prerequisite for the ergonomic design of tools that are used to manipulate objects.
Similar articles
-
Effect of laparoscopic grasper force transmission ratio on grasp control.Surg Endosc. 2009 Apr;23(4):818-24. doi: 10.1007/s00464-008-0107-6. Epub 2008 Sep 24. Surg Endosc. 2009. PMID: 18814010
-
The effects of digital anesthesia on force control using a precision grip.J Neurophysiol. 2003 Feb;89(2):672-83. doi: 10.1152/jn.00434.2001. J Neurophysiol. 2003. PMID: 12574445
-
Delayed visual feedback affects both manual tracking and grip force control when transporting a handheld object.J Neurophysiol. 2010 Aug;104(2):641-53. doi: 10.1152/jn.00174.2010. Epub 2010 Jun 10. J Neurophysiol. 2010. PMID: 20538774
-
Grip force regulation during pinch grip lifts under somatosensory guidance: comparison between people with stroke and healthy controls.Arch Phys Med Rehabil. 2006 Mar;87(3):418-29. doi: 10.1016/j.apmr.2005.11.018. Arch Phys Med Rehabil. 2006. PMID: 16500179
-
Grasp stiffness as a function of grasp force and finger span.Motor Control. 1998 Oct;2(4):352-78. doi: 10.1123/mcj.2.4.352. Motor Control. 1998. PMID: 9758886 Review.
Cited by
-
The effects of laparoscopic graspers with enhanced haptic feedback on applied forces: a randomized comparison with conventional graspers.Surg Endosc. 2017 Dec;31(12):5411-5417. doi: 10.1007/s00464-017-5623-9. Epub 2017 Jun 7. Surg Endosc. 2017. PMID: 28593415 Free PMC article. Clinical Trial.
-
Defining the Relationship Between Compressive Stress and Tissue Trauma During Laparoscopic Surgery Using Human Large Intestine.IEEE J Transl Eng Health Med. 2019 Jul 24;7:3300108. doi: 10.1109/JTEHM.2019.2919029. eCollection 2019. IEEE J Transl Eng Health Med. 2019. PMID: 31410319 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources