A biomechanical comparison of 2 transosseous-equivalent double-row rotator cuff repair techniques using bioabsorbable anchors: cyclic loading and failure behavior
- PMID: 19664507
- DOI: 10.1016/j.arthro.2009.02.023
A biomechanical comparison of 2 transosseous-equivalent double-row rotator cuff repair techniques using bioabsorbable anchors: cyclic loading and failure behavior
Abstract
Purpose: A novel double-row configuration was compared with a traditional double-row configuration for rotator cuff repair.
Methods: In 10 matched-pair sheep shoulders in vitro repair was performed with either a double-row technique with corkscrew suture anchors for the medial row and insertion anchors for the lateral row (group A) or a double-row technique with a new tape-like suture material with insertion anchors for both the medial and lateral rows (group B). Each specimen underwent cyclic loading from 10 to 150 N for 100 cycles, followed by unidirectional failure testing. Gap formation and strain within the repair area for the first and last cycles were analyzed with a video digitizing system, and stiffness and failure load were determined from the load-elongation curve.
Results: The results were similar for the 2 repair types. There was no significant difference between the ultimate failure loads of the 2 techniques (421 +/- 150 N in group A and 408 +/- 66 N in group B, P = .31) or the stiffness of the 2 techniques (84 +/- 26 N/mm in group A and 99 +/- 20 N/mm in group B, P = .07). In addition, gap formation was not different between the repair types. Strain over the repair area was also not different between the repair types.
Conclusions: Both tested rotator cuff repair techniques had high failure loads, limited gap formation, and acceptable strain patterns. No significant difference was found between the novel and conventional double-row repair types.
Clinical relevance: Two double-row techniques-one with corkscrew suture anchors for the medial row and insertion anchors for the lateral row and one with insertion anchors for both the medial and lateral rows-provided excellent biomechanical profiles at time 0 for double-row repairs in a sheep model. Although the sheep model may not directly correspond to in vivo conditions, all-insertion anchor double-row constructs are worthy of further investigation.
Similar articles
-
Cyclic loading of rotator cuff reconstructions: single-row repair with modified suture configurations versus double-row repair.Am J Sports Med. 2008 Aug;36(8):1504-10. doi: 10.1177/0363546508314424. Epub 2008 Feb 22. Am J Sports Med. 2008. PMID: 18296541
-
Biomechanical advantages of triple-loaded suture anchors compared with double-row rotator cuff repairs.Arthroscopy. 2010 Mar;26(3):316-23. doi: 10.1016/j.arthro.2009.07.019. Epub 2009 Dec 29. Arthroscopy. 2010. PMID: 20206040
-
The effect of dynamic external rotation comparing 2 footprint-restoring rotator cuff repair techniques.Am J Sports Med. 2008 May;36(5):893-900. doi: 10.1177/0363546507313092. Epub 2008 Feb 13. Am J Sports Med. 2008. PMID: 18272799
-
Arthroscopic rotator cuff repairs: an anatomic and biomechanical rationale for different suture-anchor repair configurations.Arthroscopy. 2007 Jun;23(6):662-9. doi: 10.1016/j.arthro.2007.02.018. Arthroscopy. 2007. PMID: 17560482 Review.
-
Transosseous-equivalent rotator cuff repair: a systematic review on the biomechanical importance of tying the medial row.Arthroscopy. 2013 Feb;29(2):377-86. doi: 10.1016/j.arthro.2012.11.008. Arthroscopy. 2013. PMID: 23369482
Cited by
-
Comparison of clinical outcomes between knotted and knotless double-row arthroscopic rotator cuff repairs: a meta-analysis.JSES Int. 2020 Dec 5;5(2):254-260. doi: 10.1016/j.jseint.2020.10.007. eCollection 2021 Mar. JSES Int. 2020. PMID: 33681845 Free PMC article. Review.
-
Clinical and structural evaluation of arthroscopic double-row suture-bridge rotator cuff repair: early results of a novel technique.Knee Surg Sports Traumatol Arthrosc. 2010 Dec;18(12):1730-7. doi: 10.1007/s00167-010-1257-3. Epub 2010 Sep 8. Knee Surg Sports Traumatol Arthrosc. 2010. PMID: 20824269
-
Does the Anchor Eye Alignment Affect the Performance of the Rotator Cuff Repair? A Biomechanical Study.Arthrosc Sports Med Rehabil. 2021 Feb 1;3(2):e421-e425. doi: 10.1016/j.asmr.2020.10.005. eCollection 2021 Apr. Arthrosc Sports Med Rehabil. 2021. PMID: 34027450 Free PMC article.
-
Rotator Cuff Tears at the Musculotendinous Junction: Classification and Surgical Options for Repair and Reconstruction.Arthrosc Tech. 2017 Jul 24;6(4):e1075-e1085. doi: 10.1016/j.eats.2017.03.023. eCollection 2017 Aug. Arthrosc Tech. 2017. PMID: 28970995 Free PMC article.
-
Superiority of bridging techniques with medial fixation on initial strength.Knee Surg Sports Traumatol Arthrosc. 2012 Dec;20(12):2559-66. doi: 10.1007/s00167-012-1922-9. Epub 2012 Feb 26. Knee Surg Sports Traumatol Arthrosc. 2012. PMID: 22367007
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous