Comparison of horizontal corneal diameter measurements using Galilei, EyeSys and Orbscan II systems
- PMID: 19681922
- DOI: 10.1111/j.1444-0938.2009.00407.x
Comparison of horizontal corneal diameter measurements using Galilei, EyeSys and Orbscan II systems
Abstract
Background: White-to-white corneal measurement is used frequently in cataract and refractive surgery as well as in diagnosing various corneal diseases and monitoring congenital glaucoma. In this article, we determine agreement in horizontal corneal diameter measurements in normal eyes among the Galilei V4.01, the EyeSys and the Orbscan IIz.
Methods: In this prospective study, 37 consecutive volunteers were enrolled. All 74 eyes were measured with all three devices. Galilei, EyeSys and Orbscan II measurements were compared with one another. In addition, the data from the right and left eyes were compared for each device.
Results: The mean horizontal corneal diameter measurements with the Galilei, the EyeSys and the Orbscan II were 12.01 +/- 0.61 mm, 12.09 +/- 0.87 mm and 11.67 +/- 0.29 mm, respectively. The difference between the Galilei and EyeSys systems was not statistically significant (p = 0.76), however, the Orbscan II showed smaller measurements (p < 0.001). The best 95% limits of agreement between devices were for the Galilei and the Orbscan II (-0.72, 1.48; r = 0.4, p < 0.01). The best 95% limits of agreement between two eyes for each device were found with the Orbscan II (-0.15, 0.17; r = 0.996, p < 0.01). There was no correlation between measurements of right and left eyes for the Galilei or the EyeSys (r = 0.274, p = 0.176 and r = 0.31, p = 0.065, respectively).
Conclusions: These results suggest that measurements made with the Orbscan II are smaller than those obtained with the EyeSys Corneal Analysis system and the Galilei. Among the three devices, the Galilei and the Orbscan II showed the best agreement, however, it is inadvisable to use the three devices interchangeably in every clinical situation.
Similar articles
-
Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems.Am J Ophthalmol. 2013 Jul;156(1):53-60. doi: 10.1016/j.ajo.2013.01.029. Epub 2013 Mar 28. Am J Ophthalmol. 2013. PMID: 23540708
-
Comparison of Horizontal corneal diameter measurements using the Orbscan IIz and Pentacam HR systems.Cornea. 2013 Nov;32(11):1460-4. doi: 10.1097/ICO.0b013e3182a40786. Cornea. 2013. PMID: 24055904
-
Corneal thickness measurement using Orbscan, Pentacam, Galilei, and ultrasound in normal and post-femtosecond laser in situ keratomileusis eyes.Cornea. 2012 Sep;31(9):978-82. doi: 10.1097/ICO.0b013e31823d03fc. Cornea. 2012. PMID: 22699561
-
Corneal imaging with slit-scanning and Scheimpflug imaging techniques.Clin Exp Optom. 2011 Jan;94(1):33-42. doi: 10.1111/j.1444-0938.2010.00509.x. Epub 2010 Aug 16. Clin Exp Optom. 2011. PMID: 20718786 Review.
-
Galileo Galilei: Science vs. faith.Glob Cardiol Sci Pract. 2017 Jun 30;2017(2):10. doi: 10.21542/gcsp.2017.10. Glob Cardiol Sci Pract. 2017. PMID: 29644222 Free PMC article. Review. No abstract available.
Cited by
-
Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.Int J Ophthalmol. 2016 Jul 18;9(7):1057-65. doi: 10.18240/ijo.2016.07.20. eCollection 2016. Int J Ophthalmol. 2016. PMID: 27500117 Free PMC article. Review.
-
Corneal white-to-white distance and mesopic pupil diameter.Int J Ophthalmol. 2012;5(4):505-9. doi: 10.3980/j.issn.2222-3959.2012.04.19. Epub 2012 Aug 18. Int J Ophthalmol. 2012. PMID: 22937514 Free PMC article.
-
Videographic Analysis of Blink Dynamics following Upper Eyelid Blepharoplasty and Its Association with Dry Eye.Plast Reconstr Surg Glob Open. 2020 Jul 21;8(7):e2991. doi: 10.1097/GOX.0000000000002991. eCollection 2020 Jul. Plast Reconstr Surg Glob Open. 2020. PMID: 32802679 Free PMC article.
-
Confounding sizing in posterior chamber phakic lens selection due to white-to-white measurement bias.Indian J Ophthalmol. 2019 Mar;67(3):344-349. doi: 10.4103/ijo.IJO_613_18. Indian J Ophthalmol. 2019. PMID: 30777951 Free PMC article.
-
Comparison of white-to-white measurements using four devices and their determination of ICL sizing.Eye Vis (Lond). 2022 Oct 2;9(1):36. doi: 10.1186/s40662-022-00308-z. Eye Vis (Lond). 2022. PMID: 36182955 Free PMC article.