Contrast-Enhanced Ultrasound (CEUS) for the characterization of focal liver lesions - prospective comparison in clinical practice: CEUS vs. CT (DEGUM multicenter trial). Parts of this manuscript were presented at the Ultrasound Dreiländertreffen 2008, Davos
- PMID: 19688670
- DOI: 10.1055/s-0028-1109673
Contrast-Enhanced Ultrasound (CEUS) for the characterization of focal liver lesions - prospective comparison in clinical practice: CEUS vs. CT (DEGUM multicenter trial). Parts of this manuscript were presented at the Ultrasound Dreiländertreffen 2008, Davos
Abstract
Aim: The aim of our study was to evaluate the diagnostic value of Contrast-Enhanced Ultrasound (CEUS) for the characterization of focal liver lesions in a prospective multi-center study in clinical practice. For this purpose CEUS was compared with the spiral-CT (SCT), the standard radiological method.
Material and methods: 1349 patients with unclear liver lesions after fundamental ultrasound diagnostics including color doppler analysis were examined with standardized CEUS (pulse inversion method, mechanical index < 0.4) from May 2004 to December 2006 in 14 hospitals in a prospective study. The enhancement of contrast medium in the liver tumors was analyzed according to known tumor-specific vascular patterns, using standardized documentation and analysis methods for the differentiation of tumor differentiation (malign or benign) and tumor specification (entity). A subcollective of 267 patients was additionally examined by standardized SCT method. Final diagnosis was based on histology, SCT or MRI in typical findings of liver hemangioma and FNH and on proved clinical data and additional follow up.
Results: The subcollective of 267 patients was divided in two subgroups. In 109 of these patients (subgroup A) there was no histological verification, diagnoses based on clear SCT-findings in 79 cases of hemangioma or FNH, as well as in 20 cases with a clear clinical diagnosis. 6 cases (5.5 %) remained unclear. In this subgroup the assessment of tumor differentiation was concordant with CEUS in 90 cases, discordant in 19 cases and the assessment of tumor specification was concordant in 82 and and discordant in 27 cases. In 158 patients (subgroup B) a histological finding was also present, only in 4 cases no definitive tumor diagnosis was achieved. In this subgroup assessment of tumor differentiation with CEUS and SCT was concordant in 124 cases and discordant in 30 cases (CEUS/SCT: sensitivity 94.0 / 90.7 %, specificity 83.0 / 81.5 %, PPV 91.6 / 91.5 %, NPV 87.5 / 80.0 %, accuracy 90.3 / 87.8 %). Tumor specification matched in 103cases and were different in 51 cases (CEUS/SCT: sensitivity 95.3 / 90.6 %, specificity 83.7 / 81.6 %, PPV 92.7 / 91.4 %, NPV 89.1 / 80.0 %, accuracy 91.6 / 87.7 %). A statistically significant difference could not be established. The analysis of particular tumor specification showed a statistically non significant slight advantage in tumor differentiation for CEUS in the case of hemangioma, FNH, HCC and metastases.
Conclusion: In a multi-center approach under routine clinical conditions, this prospective study demonstrates CEUS to be of equal rank to the CT-scan in regard to the assessment of tumor differentiation and specification. No statistically significant differences could be established. CEUS should be employed before computed tomography is performed for the differentiation of liver tumors, because radiation exposure and invasive biopsies can be avoided in veritable numbers of cases, when precise clinical evaluation of the findings is implemented.
(c) Georg Thieme Verlag KG Stuttgart, New York.
Similar articles
-
Contrast-enhanced ultrasound (CEUS) for the characterization of focal liver lesions in clinical practice (DEGUM Multicenter Trial): CEUS vs. MRI--a prospective comparison in 269 patients.Ultraschall Med. 2010 Oct;31(5):492-9. doi: 10.1055/s-0029-1245591. Epub 2010 Jul 22. Ultraschall Med. 2010. PMID: 20652854
-
Tumor-specific vascularization pattern of liver metastasis, hepatocellular carcinoma, hemangioma and focal nodular hyperplasia in the differential diagnosis of 1,349 liver lesions in contrast-enhanced ultrasound (CEUS).Ultraschall Med. 2009 Aug;30(4):376-82. doi: 10.1055/s-0028-1109672. Epub 2009 Aug 17. Ultraschall Med. 2009. PMID: 19688669
-
[Dynamic MRI of the liver with parallel acquisition technique: characterization of focal liver lesions and analysis of the hepatic vasculature in a single MRI session].Rofo. 2008 May;180(5):440-8. doi: 10.1055/s-2008-1027279. Rofo. 2008. PMID: 18438745 German.
-
[Focal liver lesion, incidental finding].Dtsch Med Wochenschr. 2012 Oct;137(41):2099-116. doi: 10.1055/s-0032-1305302. Epub 2012 Oct 2. Dtsch Med Wochenschr. 2012. PMID: 23033169 Review. German.
-
[Integrated diagnosis of liver angioma: comparison of Doppler color ultrasonography, computerized tomography, and magnetic resonance].Radiol Med. 1997 Jan-Feb;93(1-2):87-94. Radiol Med. 1997. PMID: 9380876 Review. Italian.
Cited by
-
[Ultrasound in oncology: screening and staging].Internist (Berl). 2012 Mar;53(3):271-81. doi: 10.1007/s00108-011-2958-5. Internist (Berl). 2012. PMID: 22354462 Review. German.
-
Evaluation of Contrast-Enhanced Intraoperative Ultrasound in the Detection and Management of Liver Lesions in Patients with Hepatocellular Carcinoma.J Oncol. 2019 Aug 5;2019:6089340. doi: 10.1155/2019/6089340. eCollection 2019. J Oncol. 2019. PMID: 31467539 Free PMC article.
-
Contrast-enhanced ultrasound is useful for the evaluation of focal liver lesions in children.Australas J Ultrasound Med. 2021 Jun 27;24(3):143-150. doi: 10.1002/ajum.12279. eCollection 2021 Aug. Australas J Ultrasound Med. 2021. PMID: 34765424 Free PMC article.
-
[Contrast-enhanced sonography. Therapy control of radiofrequency ablation and transarterial chemoembolization of hepatocellular carcinoma].Radiologe. 2011 Jun;51(6):462-8. doi: 10.1007/s00117-010-2101-y. Radiologe. 2011. PMID: 21557022 German.
-
Diagnostic value of contrast-enhanced ultrasound versus computed tomography for hepatocellular carcinoma: a retrospective, single-center evaluation of 234 patients.J Int Med Res. 2020 Jun;48(6):300060520930151. doi: 10.1177/0300060520930151. J Int Med Res. 2020. PMID: 32529869 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical