Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2009 Jun 17;10(3):37-48.
doi: 10.1120/jacmp.v10i3.3023.

Image quality & dosimetric property of an investigational imaging beam line MV-CBCT

Affiliations

Image quality & dosimetric property of an investigational imaging beam line MV-CBCT

Chris Beltran et al. J Appl Clin Med Phys. .

Abstract

To measure and compare the contrast to noise ratio (CNR) as a function of dose for the CBCTs produced by the mega-voltage (MV) imaging beam line (IBL) and the treatment beam line (TBL), and to compare the dose to target and various critical structures of pediatric patients for the IBL CBCT versus standard TBL orthogonal port films. Two Siemens Oncor linear accelerators were modified at our institution such that the MV-CBCT would operate under an investigational IBL rather than the standard 6MV TBL. Prior to the modification, several CBCTs of an electron density phantom were acquired with the TBL at various dose values. After the modification, another set of CBCTs of the electron density phantom were acquired for various doses using the IBL. The Contrast to Noise Ratio (CNR) for each tissue equivalent insert was calculated. In addition, a dosimetric study of pediatric patients was conducted comparing the 1 cGy IBL CBCT and conventional TBL orthogonal pair port films. The CNR for eight tissue equivalent inserts at five different dose settings for each type of CBCT was measured. The CNR of the muscle insert was 0.8 for a 5 cGy TBL CBCT, 1.1 for a 1.5 cGy IBL CBCT and 2.8 for a conventional CT. The CNR of the trabecular bone insert was 2.9 for a 5 cGy TBL CBCT, 5.5 for a 1.5 cGy IBL CBCT and 14.8 for a conventional CT. The IBL CBCT delivered approximately one-fourth the dose to the target and critical structures of the patients as compared to the TBL orthogonal pair port films. The IBL CBCT improves image quality while simultaneously reducing the dose to the patient as compared to the TBL CBCT. A 1 cGy IBL CBCT, which is used for boney anatomy localization, delivers one-fourth the dose as compared to conventional ortho-pair films.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Image (left) and CT (right) of the heterogeneity phantom. The regions of interests shown on the CT are the same that are used for the various CBCTs.
Figure 2
Figure 2
The center slice of the heterogeneity phantom with corresponding regions of interests for a 2 cGy IBL CBCT (left) and a 5 cGy TBL CBCT (right).
Figure 3
Figure 3
The relative dose distribution for the TBL (left) and IBL (right) CBCT. The 100% line intersects the isocenter.
Figure 4
Figure 4
Graph of the CNR for the low‐contrast medium in the phantom for the Imaging Beam Line (solid) and Treatment Beam Line (dashed) CBCTs.
Figure 5
Figure 5
Graph of the CNR for the high‐contrast medium in the phantom for the Imaging Beam Line (solid) and Treatment Beam Line (dashed) CBCTs.
Figure 6
Figure 6
The dose distributions, normalized to isocenter, on transverse, coronal and sagittal planes along the isocenter of a cranial and a body patient for an IBL CBCT and ortho‐pair plan.
Figure 7
Figure 7
The CBCT (solid) and ortho‐pair (dashed) Dose Volume Histograms (DVH) for the clinical target volume (CTV) of the 23 cranial patients.
Figure 8
Figure 8
The CBCT (solid) and ortho‐pair (dashed) Dose Volume Histograms (DVH) for the optic chiasm of the 23 cranial patients.
Figure 9
Figure 9
The CBCT (solid) and ortho‐pair (dashed) Dose Volume Histograms (DVH) for the left kidney of the 10 body patients.
Figure 10
Figure 10
An image of a 1 cGy IBL CBCT for a 4‐year‐old female patient.

References

    1. Jaffray DA, Drake DG, Moreau M, Martinez AA, Wong JW. A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft‐tissue targets. Int J Radiat Oncol Biol Phys. 1999;45(3):773–89. - PubMed
    1. White EA, Cho J, Vallis KA, et al. Cone beam computed tomography guidance for setup of patients receiving accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2007;68(2):547–54. - PubMed
    1. Purdie TG, Bissonnette JP, Franks K, et al. Cone‐beam computed tomography for on‐line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position. Int J Radiat Oncol Biol Phys. 2007;68(1):243–52. - PubMed
    1. Moseley DJ, White EA, Wiltshire KL, et al. Comparison of localization performance with implanted fiducial markers and cone‐beam computed tomography for on‐line image‐guided radiotherapy of the prostate. Int J Radiat Oncol Biol Phys. 2007;67(3):942–53. - PMC - PubMed
    1. Létourneau D, Wong R, Moseley D, et al. Online planning and delivery technique for radiotherapy of spinal metastases using cone‐beam CT: image quality and system performance. Int J Radiat Oncol Biol Phys. 2007;67(4):1229–37. - PubMed

Publication types

LinkOut - more resources