Comparative Study
. 2010 Jun;14(2):109-14.
doi: 10.1007/s11325-009-0290-2. Epub 2009 Aug 28.

Diagnostic performance of single airflow channel recording (ApneaLink) in home diagnosis of sleep apnea

Affiliations
Comparative Study

Diagnostic performance of single airflow channel recording (ApneaLink) in home diagnosis of sleep apnea

Regine Ragette et al. Sleep Breath. 2010 Jun.

Abstract

Background: ApneaLink is a novel single-channel screening device for sleep apnea detection which is based on pressure-transduced measurement of oronasal airflow, summarised as respiratory disturbance index per hour of recording time (RDI(ApneaLink)). We tested ApneaLink's diagnostic performance in a patient population with high prevalence of sleep apnea.

Methods: ApneaLink was applied simultaneously with in-laboratory polysomnography (PSG) (n = 102, 24 female, age 54.7 years) and sequentially with PSG in the unattended home setting (n = 131, 37 female, age 59.1 years). Predictive values were computed for apnea-defining thresholds of apnea-hypopnea index (AHI) >or= 5/h, >or=10/h, >or=15/h. Night-to-night variability (NNV) was assessed over three consecutive nights (n = 55, 10 female, age 48.9 years).

Results: RDI(ApneaLink) correlated well with apnea-hypopnea index on PSG (PSG(AHI)) on simultaneous (r = 0.98, bias -0.7) and unattended home application (r = 0.95, bias -0.6). Predictive values were highest at AHI >or= 10/h (in-laboratory: sensitivity 91.1%, specificity 87.5%, LR+ (positive likelihood ratio) 7.4, LR- 0.1; home: sensitivity 80%, specificity 85.5%, LR+5.5, LR- 0.2). NNV was low (V = 0.58 +/- 0.44, range 0-1.69).

Conclusion: ApneaLink is an accurate screening tool for sleep apnea in a population with high prevalence of the disorder.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Sleep. 2000 Feb 1;23(1):61-9 - PubMed
    1. Sleep. 1999 Mar 15;22(2):217-23 - PubMed
    1. Pneumologie. 2003 Dec;57(12):734-40 - PubMed
    1. Eur Respir J. 2003 Feb;21(2):253-9 - PubMed
    1. Chest. 1991 Nov;100(5):1281-6 - PubMed

Publication types

LinkOut - more resources