[Comparison of quantitative results among two automated Rapid Plasma Reagin (RPR) assays and a manual RPR test]
- PMID: 19726896
- DOI: 10.3343/kjlm.2009.29.4.331
[Comparison of quantitative results among two automated Rapid Plasma Reagin (RPR) assays and a manual RPR test]
Abstract
Background: We compared two automated Rapid Plasma Reagin (RPR) assay kits with a manual RPR assay kit to evaluate the possibility of using the two automated RPR assays as an alternative to the manual RPR assay for a quantitative monitoring.
Methods: One hundred eighty-five samples were analyzed, including 16 sera from patients with primary, secondary, and latent syphilis. Measured RPR unit (R.U.) values of two automated RPR assay kits, Mediace RPR (Sekisui Chemical Co., Ltd, Japan) and HBi Auto RPR (HBI Co., Ltd, Korea), were compared with the RPR titers of Macro-Vue RPR card test (Becton Dickinson BD Microbiology systems, USA). As a confirmatory test, Anti-Treponema pallidum EUROLINE WB (IgG) and Anti-Treponema pallidum EUROLINE WB (IgM) (Euroimmun, Germany) were used.
Results: There was a prozone effect with Mediace RPR at RPR titer (card test) of 1:16, but not with HBi Auto RPR. The R.U. values of the two automated RPR assays did not show proportional increase to the RPR titer. Agreement between manual RPR and two automated RPR assay kits, Mediace RPR assay and HBi Auto RPR assay, were 83.8% and 83.2%, respectively.
Conclusions: The two automated RPR assay kits could not be used as an alternative to manual RPR test for quantitative analysis of RPR titer. As Mediace RPR shows a prozone effect at relatively low RPR titer, caution is needed in the interpretation of the measured values.
Similar articles
-
Comparison of an automated rapid plasma reagin (RPR) test with the conventional RPR card test in syphilis testing.BMJ Open. 2014 Dec 31;4(12):e005664. doi: 10.1136/bmjopen-2014-005664. BMJ Open. 2014. PMID: 25552608 Free PMC article.
-
Usefulness of Automated Latex Turbidimetric Rapid Plasma Reagin Test for Diagnosis and Evaluation of Treatment Response in Syphilis in Comparison with Manual Card Test: a Prospective Cohort Study.J Clin Microbiol. 2018 Oct 25;56(11):e01003-18. doi: 10.1128/JCM.01003-18. Print 2018 Nov. J Clin Microbiol. 2018. PMID: 30135229 Free PMC article.
-
[Analysis of positive results in mediace rapid plasma reagin and Treponema pallidum latex agglutination as the automated syphilis test].Korean J Lab Med. 2007 Oct;27(5):324-9. doi: 10.3343/kjlm.2007.27.5.324. Korean J Lab Med. 2007. PMID: 18094596 Korean.
-
[Evaluation of performance and false positivity of Mediace RPR test that uses a chemistry autoanalyzer].Korean J Lab Med. 2008 Aug;28(4):312-8. doi: 10.3343/kjlm.2008.28.4.312. Korean J Lab Med. 2008. PMID: 18728382 Korean.
-
Evaluation of a Multiplex Fully Automated Treponemal and Nontreponemal (Rapid Plasma Reagin) Assay.Am J Clin Pathol. 2019 Jul 5;152(2):230-236. doi: 10.1093/ajcp/aqz034. Am J Clin Pathol. 2019. PMID: 31139835
Cited by
-
Comparison of Manual and Fully Automated AIX1000 Rapid Plasma Reagin Assays for Laboratory Diagnosis of Syphilis.J Clin Microbiol. 2018 Jul 26;56(8):e00214-18. doi: 10.1128/JCM.00214-18. Print 2018 Aug. J Clin Microbiol. 2018. PMID: 29618500 Free PMC article.
-
Comparison of automated treponemal and nontreponemal test algorithms as first-line syphilis screening assays.Ann Lab Med. 2016 Jan;36(1):23-7. doi: 10.3343/alm.2016.36.1.23. Ann Lab Med. 2016. PMID: 26522755 Free PMC article.
-
Comparison of an automated rapid plasma reagin (RPR) test with the conventional RPR card test in syphilis testing.BMJ Open. 2014 Dec 31;4(12):e005664. doi: 10.1136/bmjopen-2014-005664. BMJ Open. 2014. PMID: 25552608 Free PMC article.
-
Usefulness of Automated Latex Turbidimetric Rapid Plasma Reagin Test for Diagnosis and Evaluation of Treatment Response in Syphilis in Comparison with Manual Card Test: a Prospective Cohort Study.J Clin Microbiol. 2018 Oct 25;56(11):e01003-18. doi: 10.1128/JCM.01003-18. Print 2018 Nov. J Clin Microbiol. 2018. PMID: 30135229 Free PMC article.
-
Challenges of Rapid Plasma Reagin Interpretation in Syphilis Screening in Uganda: Variability in Nontreponemal Results Between Different Laboratories.Sex Transm Dis. 2018 Dec;45(12):829-833. doi: 10.1097/OLQ.0000000000000883. Sex Transm Dis. 2018. PMID: 29944643 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Medical
Research Materials