Suitability of a liquid chromatography assay of neomycin sulfate to replace the microbiological assay for neomycin in USP Monographs
- PMID: 19748755
- DOI: 10.1016/j.jpba.2009.08.011
Suitability of a liquid chromatography assay of neomycin sulfate to replace the microbiological assay for neomycin in USP Monographs
Abstract
The current USP National Formulary contains 65 Monographs for drug formulations containing neomycin. All 65 Monographs prescribe a bioassay for neomycin assay. This bioassay, based on cell culture, is labor intensive, has poor precision, and cannot be adapted for purity or identification. High-performance anion-exchange chromatography with integrated pulsed amperometric detection (HPAE-IPAD), a liquid chromatography technique, has been shown to be suitable for neomycin purity analysis and neomycin assay of an over-the-counter first aid cream (Hanko and Rohrer [17]). Here we propose that an HPAE-IPAD assay can replace the bioassay in the 65 neomycin-containing Monographs. We applied the HPAE-IPAD assay to four neomycin-containing drug products representing the four classes of formulations found in the 65 Monographs, liquid, solid, suspension, and cream. Each drug was analyzed with two chromatography systems, and on 3 separate days. For all products, HPAE-IPAD measurements were precise and accurate with respect to the label concentrations. There was also high accuracy for spike recovery of neomycin from the four drug products throughout 70-150% of the labeled concentration. These results suggest that an HPAE-IPAD assay would be an accurate assay for neomycin, and would be faster and more precise than the current bioassay.