Microvolt T-wave alternans during exercise and pacing are not comparable
- PMID: 19758980
- DOI: 10.1093/europace/eup253
Microvolt T-wave alternans during exercise and pacing are not comparable
Abstract
Aims: The absence of microvolt T-wave alternans (MTWA) identifies a group of patients who are at low risk for ventricular arrhythmia or sudden cardiac death. However, in exercised assessed MTWA, 20-40% of all test results are indeterminate. We hypothesised that MTWA during pacing would yield less indeterminate results.
Methods and results: Thirty patients with ischaemic cardiomyopathy and prior dual chamber implantable cardioverter defibrillator implantation were enrolled. All patients underwent sequential MTWA testing using an exercise (E), atrial-paced (A), and atrioventricular-paced (AV) protocol. The number of indeterminate tests was lower during pacing (A: 17%; AV: 3%) compared with exercise (37%) (E vs. A: P = 0.015, E vs. AV: P = <0.001). When positive and indeterminate test results were grouped as non-negative, the concordance rates between E and A, E and AV, and A and AV were 60% (kappa = 0.17), 57% (kappa = 0.058), and 70% (kappa = 0.348), respectively. If indeterminate results were excluded, agreements were 60% (kappa = 0.19), 50% (kappa = 0.129) and 67% (kappa = 0.33), respectively.
Conclusion: Indeterminate test results are less common during pacing. However, there is a low concordance rate between test results using different protocols. This necessitates further study to determine the predictive value of each method in high risk patients with ischaemic cardiomyopathy.
Comment in
-
Detection of T-wave alternans by exercise stress testing, atrial-pacing, and atrioventricular pacing.Europace. 2010 Jan;12(1):146; author reply 146-7. doi: 10.1093/europace/eup354. Europace. 2010. PMID: 20008497 No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical