. 2009 Nov;19(11):2576-83.
doi: 10.1007/s00330-009-1558-4.

Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience

Affiliations

Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience

Michael Lell et al. Eur Radiol. 2009 Nov.

Abstract

Objective: We evaluated radiation exposure and image quality of a new coronary CT angiography protocol, high-pitch spiral acquisition, using dual source CT (DSCT).

Material and methods: Coronary CTAwas performed in 25 consecutive patients with a stable heart rate of 60 bpm or less after premedication, using 2 x 128 0.6-mm sections, 38.4-mm collimation width and 0.28-s rotation time. Tube settings were 100 kV/320 mAs and 120 kV/400 mAs for patients below and above 100-kg weight, respectively. Data acquisition was prospectively ECG-triggered at 60% of the R-R interval using a pitch of 3.2 (3.4 for the last 10 patients). Images were reconstructed with 75-ms temporal resolution, 0.6-mm slice thickness and 0.3-mm increment. Image quality was evaluated using a four-point scale (1 = excellent, 4 = unevaluable).

Results: Mean range of data acquisition was 113 +/- 22 mm, mean duration was 268 +/- 23 ms. Of 363 coronary artery segments, 327 had an image quality score of 1, and only 2 segments were rated as "unevaluable". Mean dose-length product (DLP) was 71 +/- 23 mGy cm, mean effective dose was 1.0 +/- 0.3 mSv (range 0.78-2.1 mSv). For 21 patients with a body weight below 100 kg, mean DLP was 63 +/- 5 mGy cm (0.88 +/- 0.07 mSv; range 0.78-0.97 mSv).

Conclusion: Prospectively ECGtriggered high-pitch spiral CT acquisition provides high and stable image quality at very low radiation dose.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Med Phys. 2009 Mar;36(3):993-1007 - PubMed
    1. J Cardiovasc Comput Tomogr. 2008 Sep-Oct;2(5):325-7 - PubMed
    1. AJR Am J Roentgenol. 2009 Apr;192(4):1045-50 - PubMed
    1. J Am Coll Cardiol. 2007 Dec 18;50(25):2393-8 - PubMed
    1. Radiology. 2008 Mar;246(3):742-53 - PubMed

Publication types