Postoperative changes in intraocular pressure and corneal biomechanical metrics Laser in situ keratomileusis versus laser-assisted subepithelial keratectomy
- PMID: 19781475
- DOI: 10.1016/j.jcrs.2009.05.041
Postoperative changes in intraocular pressure and corneal biomechanical metrics Laser in situ keratomileusis versus laser-assisted subepithelial keratectomy
Abstract
Purpose: To compare intraocular pressure (IOP) and corneal biomechanical metric changes after myopic laser in situ keratomileusis and laser-assisted subepithelial keratectomy (LASEK).
Setting: Private practice, St. Louis, Missouri, USA.
Methods: The IOP, corneal biomechanical markers, and Ocular Response Analyzer (ORA) waveform parameters were prospectively measured preoperatively and after 6 months in ablation-matched myopic LASIK eyes (mLASIK group) and LASEK eyes (mLASEK group). A retrospectively identified cohort of low myopia LASIK eyes (lmLASIK group) and fellow unoperated eyes (control) were tested at a single postoperative visit. Statistical analysis compared the percentage change in parameters between groups.
Results: The mean postoperative Goldmann tonometry and Goldmann-correlated IOPs were statistically significant reduced in the mLASIK and mLASEK groups (P<.03). Corneal-compensated IOP, but not Pascal dynamic contour tonometry, was significantly reduced in the mLASIK group. The percentage change in corneal hysteresis (CH) and the corneal resistance factor (CRF) was greater in the mLASIK and mLASEK groups than in the lmLASIK group. The greatest percentage change in ORA signal parameters was in the mLASIK group and the smallest change, in the mLASEK group. On multivariate linear regression, the residual stromal bed was predictive of the percentage change in CH and CRF (P<.001).
Conclusions: Microkeratome flap creation combined with deeper stromal ablation had the greatest effect on the ORA applanation signal, indicating corneas that are more readily deformable. The smallest change in the signal was in the group without a stromal flap (LASEK). There was a complex interaction between ablation location and depth that affected corneal biomechanical properties.
Similar articles
-
Corneal biomechanical measurements before and after laser in situ keratomileusis.J Cataract Refract Surg. 2008 Nov;34(11):1886-91. doi: 10.1016/j.jcrs.2008.06.035. J Cataract Refract Surg. 2008. PMID: 19006734
-
Changes in biomechanical properties of the cornea and intraocular pressure after myopic laser in situ keratomileusis using a femtosecond laser for flap creation determined using ocular response analyzer and Goldmann applanation tonometry.J Glaucoma. 2015 Mar;24(3):195-201. doi: 10.1097/IJG.0b013e31829da1ec. J Glaucoma. 2015. PMID: 23807345
-
Differences in the corneal biomechanical effects of surface ablation compared with laser in situ keratomileusis using a microkeratome or femtosecond laser.J Cataract Refract Surg. 2008 Dec;34(12):2049-56. doi: 10.1016/j.jcrs.2008.08.021. J Cataract Refract Surg. 2008. PMID: 19027558
-
Corneal biomechanical properties after SMILE versus FLEX, LASIK, LASEK, or PRK: a systematic review and meta-analysis.BMC Ophthalmol. 2019 Aug 1;19(1):167. doi: 10.1186/s12886-019-1165-3. BMC Ophthalmol. 2019. PMID: 31370817 Free PMC article.
-
Thin-flap laser-assisted in situ keratomileusis.Curr Opin Ophthalmol. 2008 Jul;19(4):325-9. doi: 10.1097/ICU.0b013e328302cc77. Curr Opin Ophthalmol. 2008. PMID: 18545016 Review.
Cited by
-
Changes in custom biomechanical variables after femtosecond laser in situ keratomileusis and photorefractive keratectomy for myopia.J Cataract Refract Surg. 2014 Jun;40(6):918-28. doi: 10.1016/j.jcrs.2013.11.030. Epub 2014 Apr 13. J Cataract Refract Surg. 2014. PMID: 24726160 Free PMC article.
-
Short-term effects of overnight orthokeratology on corneal epithelial permeability and biomechanical properties.Invest Ophthalmol Vis Sci. 2013 Jun 6;54(6):3902-11. doi: 10.1167/iovs.13-11874. Invest Ophthalmol Vis Sci. 2013. PMID: 23652492 Free PMC article.
-
Corneal Biomechanics in Ectatic Diseases: Refractive Surgery Implications.Open Ophthalmol J. 2017 Jul 31;11:176-193. doi: 10.2174/1874364101711010176. eCollection 2017. Open Ophthalmol J. 2017. PMID: 28932334 Free PMC article.
-
Comparison of intraocular pressure measurements obtained by Goldmann applanation tonometer, corvis ST and a conventional non-contact airpuff tonometer in eyes with previous myopic refractive surgery and correlation with corneal biomechanical parameters.Int Ophthalmol. 2025 Jun 6;45(1):232. doi: 10.1007/s10792-025-03598-z. Int Ophthalmol. 2025. PMID: 40478442
-
Visual outcomes after three different surgical procedures for correction of refractive error in patients with thin corneas.Int J Ophthalmol. 2020 Jun 18;13(6):970-977. doi: 10.18240/ijo.2020.06.18. eCollection 2020. Int J Ophthalmol. 2020. PMID: 32566511 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources