Inaccuracy of the Hawksley random zero sphygmomanometer
Abstract
To examine the accuracy of the Hawksley random zero sphygmomanometer two studies were done with subjects with a wide range of blood pressure. When readings made by one observer on the UK model of the Hawksley sphygmomanometer were compared with readings by two independent observers on separate mercury sphygmomanometers, the Hawksley device underestimated systolic readings by a mean (SD) of 2.0 (2.4) and 0.5 (3.6) mm Hg and diastolic readings by a mean of 3.7 (2.7) and 2.8 (2.9) mm Hg. When readings made on the UK and US models of the Hawksley sphygmomanometer were compared with those made on mercury sphygmomanometers, with observers exchanging devices half way during the experiment, the UK Hawksley device underestimated systolic pressure by a mean of 3.8 (SD 3.5) mm Hg and diastolic blood pressure by 7.5 (3.8) mm Hg; and the US model by 2.6 (3.4) mm Hg for systolic pressure and 6.2 (3.7) mm Hg for diastolic pressure. There was better agreement between two observers using standard sphygmomanometers than between an observer using the Hawksley random zero sphygmomanometer and an observer using a standard sphygmomanometer. Thus, the quantitative aspects of blood pressure in epidemiological and intervention studies in which the Hawksley random zero sphygmomanometer was used need re-evaluation. Moreover, the Hawksley random zero sphygmomanometer, in its present design, should not be used in hypertension research.
Comment in
-
Inaccuracy of the Hawksley random zero sphygmomanometer.Lancet. 1991 Mar 2;337(8740):556. doi: 10.1016/0140-6736(91)91344-t. Lancet. 1991. PMID: 1671917 No abstract available.
-
Inaccuracy of the Hawksley random-zero sphygmomanometer.Lancet. 1991 Apr 6;337(8745):866. Lancet. 1991. PMID: 1672964 No abstract available.
Similar articles
-
Does the Hawksley random zero sphygmomanometer underestimate blood pressure, and by how much?J Hum Hypertens. 1995 May;9(5):337-43. J Hum Hypertens. 1995. PMID: 7623370 Clinical Trial.
-
Using Hawksley random zero sphygmomanometer as a gold standard may result in misleading conclusions.Blood Press. 1994 Sep;3(5):283-6. doi: 10.3109/08037059409102275. Blood Press. 1994. PMID: 7866591
-
Agreement of blood pressure measurements between random-zero and standard mercury sphygmomanometers.Am J Med Sci. 2008 Nov;336(5):373-8. doi: 10.1097/MAJ.0b013e31816956ac. Am J Med Sci. 2008. PMID: 19011391 Free PMC article.
-
Sphygmomanometer for Invasive Blood Pressure Monitoring in a Medical Mission.Anesthesiology. 2019 Feb;130(2):312. doi: 10.1097/ALN.0000000000002469. Anesthesiology. 2019. PMID: 30260896 Review. No abstract available.
-
Blood pressure measurement in childhood epidemiological studies.Circulation. 1995 Aug 15;92(4):1049-57. doi: 10.1161/01.cir.92.4.1049. Circulation. 1995. PMID: 7641339 Review.
Cited by
-
Caffeine restriction: effect on mild hypertension.BMJ. 1991 Nov 16;303(6812):1235-8. doi: 10.1136/bmj.303.6812.1235. BMJ. 1991. PMID: 1747643 Free PMC article. Clinical Trial.
-
Calibration of blood pressure measurements in the Jackson Heart Study.Blood Press Monit. 2019 Jun;24(3):130-136. doi: 10.1097/MBP.0000000000000379. Blood Press Monit. 2019. PMID: 30998553 Free PMC article.
-
The pursuit of accurate blood pressure measurement: A 35-year travail.J Clin Hypertens (Greenwich). 2017 Aug;19(8):746-752. doi: 10.1111/jch.13005. Epub 2017 Apr 21. J Clin Hypertens (Greenwich). 2017. PMID: 28429879 Free PMC article. No abstract available.
-
Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association.Hypertension. 2019 May;73(5):e35-e66. doi: 10.1161/HYP.0000000000000087. Hypertension. 2019. PMID: 30827125 Free PMC article. Review.
-
Blood pressure measurement in pregnancy: the effect of arm circumference and sphygmomanometer cuff size.Obstet Med. 2009 Sep;2(3):116-20. doi: 10.1258/om.2009.090017. Epub 2009 Sep 1. Obstet Med. 2009. PMID: 27582825 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources