The relationship between daily atrial tachyarrhythmia burden from implantable device diagnostics and stroke risk: the TRENDS study
- PMID: 19843914
- DOI: 10.1161/CIRCEP.109.849638
The relationship between daily atrial tachyarrhythmia burden from implantable device diagnostics and stroke risk: the TRENDS study
Abstract
Background: It is unknown if brief episodes of device-detected atrial fibrillation (AF) increase thromboembolic event (TE) risk.
Methods and results: TRENDS was a prospective, observational study enrolling patients with > or = 1 stroke risk factor (heart failure, hypertension, age > or = 65 years, diabetes, or prior TE) receiving pacemakers or defibrillators that monitor atrial tachycardia (AT)/AF burden (defined as the longest total AT/AF duration on any given day during the prior 30-day period). This time-varying exposure was updated daily during follow-up and related to TE risk. Annualized TE rates were determined according to AT/AF burden subsets: zero, low (<5.5 hours [median duration of subsets with nonzero burden]), and high (> or = 5.5 hours). A multivariate Cox model provided hazard ratios including terms for stroke risk factors and time-varying AT/AF burden and antithrombotic therapy. Patients (n=2486) had at least 30 days of device data for analysis. During a mean follow-up of 1.4 years, annualized TE risk (including transient ischemic attacks) was 1.1% for zero, 1.1% for low, and 2.4% for high burden subsets of 30-day windows. Compared with zero burden, adjusted hazard ratios (95% CIs) in the low and high burden subsets were 0.98 (0.34 to 2.82, P=0.97) and 2.20 (0.96 to 5.05, P=0.06), respectively.
Conclusions: The TE rate was low compared with patients with traditional AF with similar risk profiles. The data suggest that TE risk is a quantitative function of AT/AF burden. AT/AF burden > or = 5.5 hours on any of 30 prior days appeared to double TE risk. Additional studies are needed to more precisely investigate the relationship between stroke risk and AT/AF burden.
Trial registration: ClinicalTrials.gov NCT00279981.