Implementation of a new method for dynamic multileaf collimator tracking of prostate motion in arc radiotherapy using a single kV imager
- PMID: 19910138
- DOI: 10.1016/j.ijrobp.2009.06.073
Implementation of a new method for dynamic multileaf collimator tracking of prostate motion in arc radiotherapy using a single kV imager
Abstract
Purpose: To implement a method for real-time prostate motion estimation with a single kV imager during arc radiotherapy and to integrate it with dynamic multileaf collimator (DMLC) target tracking.
Methods and materials: An arc field with a circular aperture and 358 degrees gantry rotation was delivered to a motion phantom with a fiducial marker under continuous kV X-ray imaging at 5 Hz, perpendicular to the treatment beam. A pretreatment gantry rotation of 120 degrees in 20 sec with continuous imaging preceded the treatment. During treatment, each kV image was first used together with all previous images to estimate the three-dimensional (3D) target probability density function and then used together with this probability density function to estimate the 3D target position. The MLC aperture was then adapted to the estimated 3D target position. Tracking was performed with five patient-measured prostate trajectories that represented characteristic prostate motion patterns. Two data sets were recorded during tracking: (1) the estimated 3D target positions, for off-line comparison with the actual phantom motion; and (2) continuous portal images, for independent off-line calculation of the 2D tracking error as the positional difference between the marker and the MLC aperture center in each portal image. All experiments were also made with 1- Hz kV imaging.
Results: The mean 3D root-mean-square error of the trajectory estimation was 0.6 mm. The mean root-mean-square tracking error was 0.7 mm, both parallel and perpendicular to the MLC. The accuracy degraded slightly for 1- Hz imaging.
Conclusions: Single-imager DMLC prostate tracking that allows arbitrary beam modulation during arc radiotherapy was implemented. It has submillimeter accuracy for most prostate motion types.
Copyright (c) 2010 Elsevier Inc. All rights reserved.
Similar articles
-
Toward submillimeter accuracy in the management of intrafraction motion: the integration of real-time internal position monitoring and multileaf collimator target tracking.Int J Radiat Oncol Biol Phys. 2009 Jun 1;74(2):575-82. doi: 10.1016/j.ijrobp.2008.12.057. Epub 2009 Mar 26. Int J Radiat Oncol Biol Phys. 2009. PMID: 19327907
-
Dynamic MLC tracking of moving targets with a single kV imager for 3D conformal and IMRT treatments.Acta Oncol. 2010 Oct;49(7):1092-100. doi: 10.3109/0284186X.2010.498438. Acta Oncol. 2010. PMID: 20831501
-
Integration of real-time internal electromagnetic position monitoring coupled with dynamic multileaf collimator tracking: an intensity-modulated radiation therapy feasibility study.Int J Radiat Oncol Biol Phys. 2009 Jul 1;74(3):868-75. doi: 10.1016/j.ijrobp.2009.01.031. Epub 2009 Apr 23. Int J Radiat Oncol Biol Phys. 2009. PMID: 19394159
-
Dynamic targeting image-guided radiotherapy.Med Dosim. 2006 Summer;31(2):113-25. doi: 10.1016/j.meddos.2005.12.014. Med Dosim. 2006. PMID: 16690452 Review.
-
[A review of progress of real-time tumor tracking radiotherapy technology based on dynamic multi-leaf collimator].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2017 Feb;34(1):145-9. doi: 10.7507/1001-5515.201607054. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2017. PMID: 29717603 Free PMC article. Review. Chinese.
Cited by
-
Measurement of patient imaging dose for real-time kilovoltage x-ray intrafraction tumour position monitoring in prostate patients.Phys Med Biol. 2012 May 21;57(10):2969-80. doi: 10.1088/0031-9155/57/10/2969. Epub 2012 Apr 20. Phys Med Biol. 2012. PMID: 22517054 Free PMC article.
-
A method for robust segmentation of arbitrarily shaped radiopaque structures in cone-beam CT projections.Med Phys. 2011 Apr;38(4):2151-6. doi: 10.1118/1.3555295. Med Phys. 2011. PMID: 21626948 Free PMC article.
-
Advances in radiotherapy: from 2D to 4D.Cancer Imaging. 2011 Oct 3;11 Spec No A(1A):S147-52. doi: 10.1102/1470-7330.2011.9036. Cancer Imaging. 2011. PMID: 22185864 Free PMC article. Review.
-
Experimental investigation of a general real-time 3D target localization method using sequential kV imaging combined with respiratory monitoring.Phys Med Biol. 2012 Nov 21;57(22):7395-407. doi: 10.1088/0031-9155/57/22/7395. Epub 2012 Oct 24. Phys Med Biol. 2012. PMID: 23093356 Free PMC article.
-
Kilovoltage intrafraction monitoring for prostate intensity modulated arc therapy: first clinical results.Int J Radiat Oncol Biol Phys. 2012 Dec 1;84(5):e655-61. doi: 10.1016/j.ijrobp.2012.07.2367. Epub 2012 Sep 11. Int J Radiat Oncol Biol Phys. 2012. PMID: 22975613 Free PMC article.