Aims: The aim of this study was to evaluate the current short-term (<3 months) complication rate related to cardiac rhythm management (CRM) device implantations.
Methods and results: We analysed data of the complications related to all CRM device implantations during 1 year (2006) in a tertiary referral university hospital. In 567 device implantations, pacing system upgrade procedures, or lead revisions, 78 complications occurred in 69 (12.2%) patients. Lead dislodgement, pocket haematoma or bleeding, pneumothorax, and infection were the most common accounting for >80% of all complications. The complication rate was more than twice as high in bradycardia pacemaker (PM) implantations performed by cardiology trainees (17.4%) than by experienced cardiologists (7.7%, P = 0.001). When performed by experienced cardiologists, the complication rate was not higher in implantations of more complex devices compared with that of bradycardia PMs. Fifty-two of the 69 patients needed additional surgical procedures. Altogether, the complications required 504 additional treatment days in hospital.
Conclusion: In conclusion, our retrospective 1-year single-centre survey shows that short-term implantation-related complications of contemporary device therapy are still frequent, occur much more frequently by trainees than by cardiologists, require a large number of additional surgical procedures, and substantially prolong the hospital stay.
Comment in
-
Device implantation: the human factor.Europace. 2010 Jan;12(1):15-6. doi: 10.1093/europace/eup409. Europace. 2010. PMID: 20008495 No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical