Endocavitary thermal therapy by MRI-guided phased-array contact ultrasound: experimental and numerical studies on the multi-input single-output PID temperature controller's convergence and stability
- PMID: 19928104
- DOI: 10.1118/1.3215534
Endocavitary thermal therapy by MRI-guided phased-array contact ultrasound: experimental and numerical studies on the multi-input single-output PID temperature controller's convergence and stability
Abstract
Purpose: Endocavitary high intensity contact ultrasound (HICU) may offer interesting therapeutic potential for fighting localized cancer in esophageal or rectal wall. On-line MR guidance of the thermotherapy permits both excellent targeting of the pathological volume and accurate preoperatory monitoring of the temperature elevation. In this article, the authors address the issue of the automatic temperature control for endocavitary phased-array HICU and propose a tailor-made thermal model for this specific application. The convergence and stability of the feedback loop were investigated against tuning errors in the controller's parameters and against input noise, through ex vivo experimental studies and through numerical simulations in which nonlinear response of tissue was considered as expected in vivo.
Methods: An MR-compatible, 64-element, cooled-tip, endorectal cylindrical phased-array applicator of contact ultrasound was integrated with fast MR thermometry to provide automatic feedback control of the temperature evolution. An appropriate phase law was applied per set of eight adjacent transducers to generate a quasiplanar wave, or a slightly convergent one (over the circular dimension). A 2D physical model, compatible with on-line numerical implementation, took into account (1) the ultrasound-mediated energy deposition, (2) the heat diffusion in tissue, and (3) the heat sink effect in the tissue adjacent to the tip-cooling balloon. This linear model was coupled to a PID compensation algorithm to obtain a multi-input single-output static-tuning temperature controller. Either the temperature at one static point in space (situated on the symmetry axis of the beam) or the maximum temperature in a user-defined ROI was tracked according to a predefined target curve. The convergence domain in the space of controller's parameters was experimentally explored ex vivo. The behavior of the static-tuning PID controller was numerically simulated based on a discrete-time iterative solution of the bioheat transfer equation in 3D and considering temperature-dependent ultrasound absorption and blood perfusion.
Results: The intrinsic accuracy of the implemented controller was approximately 1% in ex vivo trials when providing correct estimates for energy deposition and heat diffusivity. Moreover, the feedback loop demonstrated excellent convergence and stability over a wide range of the controller's parameters, deliberately set to erroneous values. In the extreme case of strong underestimation of the ultrasound energy deposition in tissue, the temperature tracking curve alone, at the initial stage of the MR-controlled HICU treatment, was not a sufficient indicator for a globally stable behavior of the feedback loop. Our simulations predicted that the controller would be able to compensate for tissue perfusion and for temperature-dependent ultrasound absorption, although these effects were not included in the controller's equation. The explicit pattern of acoustic field was not required as input information for the controller, avoiding time-consuming numerical operations.
Conclusions: The study demonstrated the potential advantages of PID-based automatic temperature control adapted to phased-array MR-guided HICU therapy. Further studies will address the integration of this ultrasound device with a miniature RF coil for high resolution MRI and, subsequently, the experimental behavior of the controller in vivo.
Similar articles
-
Endoluminal MR-guided ultrasonic applicator embedding cylindrical phased-array transducers and opposed-solenoid detection coil.Magn Reson Med. 2015 Jan;73(1):417-26. doi: 10.1002/mrm.25099. Epub 2014 Jan 29. Magn Reson Med. 2015. PMID: 24478117
-
64-element intraluminal ultrasound cylindrical phased array for transesophageal thermal ablation under fast MR temperature mapping: an ex vivo study.Med Phys. 2006 Aug;33(8):2926-34. doi: 10.1118/1.2218064. Med Phys. 2006. PMID: 16964871 Free PMC article.
-
Automatic temperature control for MR-guided interstitial ultrasound ablation in liver using a percutaneous applicator: ex vivo and in vivo initial studies.Magn Reson Med. 2010 Mar;63(3):667-79. doi: 10.1002/mrm.22328. Magn Reson Med. 2010. PMID: 20187177
-
Catheter-based ultrasound applicators for selective thermal ablation: progress towards MRI-guided applications in prostate.Int J Hyperthermia. 2004 Nov;20(7):739-56. doi: 10.1080/02656730410001721816. Int J Hyperthermia. 2004. PMID: 15675669 Review.
-
Image-guided ultrasound phased arrays are a disruptive technology for non-invasive therapy.Phys Med Biol. 2016 Sep 7;61(17):R206-48. doi: 10.1088/0031-9155/61/17/R206. Epub 2016 Aug 5. Phys Med Biol. 2016. PMID: 27494561 Free PMC article. Review.
Cited by
-
Mild hyperthermia by MR-guided focused ultrasound in an ex vivo model of osteolytic bone tumour: optimization of the spatio-temporal control of the delivered temperature.J Transl Med. 2019 Oct 24;17(1):350. doi: 10.1186/s12967-019-2094-x. J Transl Med. 2019. PMID: 31651311 Free PMC article.
-
Real-time automatic temperature regulation during in vivo MRI-guided laser-induced thermotherapy (MR-LITT).Sci Rep. 2023 Feb 25;13(1):3279. doi: 10.1038/s41598-023-29818-z. Sci Rep. 2023. PMID: 36841878 Free PMC article.
-
Development of robust/predictive control strategies for image-guided ablative treatments using a minimally invasive ultrasound applicator.Int J Hyperthermia. 2014 Nov;30(7):438-46. doi: 10.3109/02656736.2014.963702. Epub 2014 Oct 14. Int J Hyperthermia. 2014. PMID: 25314227 Free PMC article.
-
A nutraceutical formulation combined with sclerofoam-assisted laser treatment ameliorates chronic venous insufficiency.Lasers Med Sci. 2022 Sep;37(7):2831-2835. doi: 10.1007/s10103-022-03549-5. Epub 2022 Apr 18. Lasers Med Sci. 2022. PMID: 35435595 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources