Ventricular actuation improves systolic and diastolic myocardial function in the small failing heart
- PMID: 19932272
- DOI: 10.1016/j.athoracsur.2009.08.048
Ventricular actuation improves systolic and diastolic myocardial function in the small failing heart
Abstract
Background: Direct mechanical ventricular actuation (DMVA) provides non-blood contacting augmentation of ventricular function. The device has promise for supporting the pediatric heart. The purpose of this study was to assess DMVA's effect in a small animal model of heart failure.
Methods: Anesthetized rabbits (n = 6) underwent sternotomy and were instrumented for hemodynamic monitoring. A 10-MHz ultrasound probe was used for transesophageal echocardiography imaging. Heart failure (cardiac output <50% baseline) was induced with esmolol. Phenylephrine was titrated to maintain baseline mean arterial pressure. Transesophageal echocardiography imaging was acquired at baseline, heart failure, and subsequent DMVA support for 2 hours. Image analysis was used to derive ejection fraction, cardiac output, and stroke work as measures of left ventricular function. Speckle tracking software was used to derive myocardial strain rates as load-independent measures of left ventricular myocardial function.
Results: Mean ejection fraction was significantly increased during DMVA support (0.585 +/- 0.035) versus failure (0.215 +/- 0.014; p < 0.001). Peak global left ventricular systolic and diastolic strain rates (1/second) were significantly increased during DMVA (-2.85 +/- 0.33 and 2.92 +/- 0.37) versus failure (-1.69 +/- 0.11 and 1.99 +/- 0.14; p < 0.001 and 0.004, respectively). Peak strain rates during DMVA in the failing heart were similar to baseline.
Conclusions: Direct mechanical ventricular actuation augments both systolic and diastolic left ventricular pump function. Diastolic augmentation distinguishes the device from other direct cardiac compression methods. This study demonstrated that DMVA in the small-sized, failing heart improves both systolic and diastolic myocardial function, which has favorable implications for left ventricular recovery. Direct mechanical ventricular actuation's salutary effects can be provided to the failing pediatric heart without complications of blood contact.
Similar articles
-
Cardiac function after acute support with direct mechanical ventricular actuation in chronic heart failure.ASAIO J. 2014 Nov-Dec;60(6):701-6. doi: 10.1097/MAT.0000000000000147. ASAIO J. 2014. PMID: 25238499
-
Autosynchronized systolic unloading during left ventricular assist with a centrifugal pump.J Thorac Cardiovasc Surg. 2003 Feb;125(2):353-60. doi: 10.1067/mtc.2003.100. J Thorac Cardiovasc Surg. 2003. PMID: 12579105
-
Comparison of acute changes in left ventricular volume, systolic and diastolic functions, and intraventricular synchronicity after biventricular and right ventricular pacing for heart failure.Am Heart J. 2003 May;145(5):E18. doi: 10.1016/S0002-8703(03)00071-1. Am Heart J. 2003. PMID: 12766742 Clinical Trial.
-
Anatomy and physiology of the right ventricle.Cardiol Clin. 2012 May;30(2):167-87. doi: 10.1016/j.ccl.2012.03.009. Cardiol Clin. 2012. PMID: 22548810 Review.
-
The forgotten driving forces in right heart failure: new concept and device.Asian Cardiovasc Thorac Ann. 2009 Oct;17(5):525-30. doi: 10.1177/0218492309348638. Asian Cardiovasc Thorac Ann. 2009. PMID: 19917800 Review.
Cited by
-
Dynamic Augmentation of Left Ventricle and Mitral Valve Function With an Implantable Soft Robotic Device.JACC Basic Transl Sci. 2020 Feb 26;5(3):229-242. doi: 10.1016/j.jacbts.2019.12.001. eCollection 2020 Mar. JACC Basic Transl Sci. 2020. PMID: 32215347 Free PMC article.
-
Human Applications of the Anstadt Cup: Implications for Non-Blood-Contacting Biventricular Support.Ann Thorac Surg Short Rep. 2024 Jun 13;2(4):842-847. doi: 10.1016/j.atssr.2024.05.019. eCollection 2024 Dec. Ann Thorac Surg Short Rep. 2024. PMID: 39790597 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical