"Double wire" angio-seal closure technique after balloon aortic valvuloplasty
- PMID: 19937771
- DOI: 10.1002/ccd.22295
"Double wire" angio-seal closure technique after balloon aortic valvuloplasty
Abstract
Objectives: To report the feasibility of a collagen-mediated closure device using a modified Angio-Seal closure technique for access site management following percutaneous balloon aortic valvuloplasty (BAV).
Background: With the advent of percutaneous aortic valve replacement therapies, there has been a resurgence of interest in BAV procedures. Vascular complications, including bleeding, are a common source of morbidity post procedure as a result of the requirement for large bore femoral artery access. The use of vascular closure devices may reduce bleeding complications.
Methods: We describe a new technique for vascular closure in this setting. At the conclusion of the valvuloplasty procedure, two 0.035'' wires are inserted through the femoral artery sheath. A conventional collagen-mediated closure device (8F Angio-Seal) is deployed over the first wire and along side the second wire. If immediate hemostasis is not achieved, a second device is loaded onto the second wire and deployed to achieve hemostasis.
Results: Percutaneous BAV was performed in 21 patients. Hemostasis was successfully achieved in all patients with either a single 8F Angio-Seal closure device (18 patients) or after placement of a second device (three patients).
Conclusions: The modified "Double Wire" Angio-Seal technique is a feasible method for hemostasis following percutaneous BAV.
Comment in
-
Improvements in balloon aortic valvuloplasty: closure devices for large artery access.Catheter Cardiovasc Interv. 2010 Mar 1;75(4):493. doi: 10.1002/ccd.22481. Catheter Cardiovasc Interv. 2010. PMID: 20333667 No abstract available.
Similar articles
-
Comparison of closure strategies after balloon aortic valvuloplasty: suture mediated versus collagen based versus manual.Catheter Cardiovasc Interv. 2011 Jul 1;78(1):119-24. doi: 10.1002/ccd.22940. Catheter Cardiovasc Interv. 2011. PMID: 21681898
-
Large calibre arterial access device closure for percutaneous aortic valve interventions: use of the Prostar system in 118 cases.Catheter Cardiovasc Interv. 2012 Jan 1;79(1):143-9. doi: 10.1002/ccd.23105. Epub 2011 Dec 8. Catheter Cardiovasc Interv. 2012. PMID: 21538787
-
Improvements in balloon aortic valvuloplasty: closure devices for large artery access.Catheter Cardiovasc Interv. 2010 Mar 1;75(4):493. doi: 10.1002/ccd.22481. Catheter Cardiovasc Interv. 2010. PMID: 20333667 No abstract available.
-
Vascular occlusion as a late complication of the Angio-Seal closure device. A review of literature.Minerva Cardioangiol. 2007 Dec;55(6):815-9. Minerva Cardioangiol. 2007. PMID: 18091648 Review.
-
Arterial thrombosis after using Angio-Seal.Ann Vasc Surg. 2011 Nov;25(8):1078-93. doi: 10.1016/j.avsg.2010.11.021. Ann Vasc Surg. 2011. PMID: 22023942 Review.
Cited by
-
Limitations of closing percutaneous transthoracic ventricular access ports using a commercial collagen vascular closure device.Catheter Cardiovasc Interv. 2011 Jun 1;77(7):1079-85. doi: 10.1002/ccd.22941. Epub 2011 Mar 21. Catheter Cardiovasc Interv. 2011. PMID: 21234923 Free PMC article.
-
Hybrid method of large bore arterial access closure: Single-center initial experience based on percutaneous coronary artery interventions assisted with left ventricle support device.Cardiol J. 2022;29(6):1040-1042. doi: 10.5603/CJ.a2022.0096. Epub 2022 Oct 6. Cardiol J. 2022. PMID: 36200543 Free PMC article. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical