Comparative Study
. 2009 Nov;27(9):335-41.
doi: 10.1007/s11604-009-0349-9. Epub 2009 Nov 27.

Spatial factors for quantifying constant flow velocity in a small tube phantom: comparison of phase-contrast cine-magnetic resonance imaging and the intraluminal Doppler guidewire method

Affiliations
Comparative Study

Spatial factors for quantifying constant flow velocity in a small tube phantom: comparison of phase-contrast cine-magnetic resonance imaging and the intraluminal Doppler guidewire method

Haruhiko Machida et al. Jpn J Radiol. 2009 Nov.

Abstract

Purpose: We examined the spatial factors influencing magnetic resonance (MR) flow velocity measurements in a small tube phantom and used the same measurements obtained with an intraluminal Doppler guidewire as reference.

Materials and methods: We generated constant flow velocities from approximately 40 to 370 cm/s in a tube 4 mm in diameter. We then performed segmented k-space, phase-contrast cine-MR imaging to quantify spatial peak flow velocities of one pixel and of five adjacent pixels as well as spatial mean velocities within regions of interest in a cross section of the phantom. Pixel dimensions ranged from 1.00 x 1.00 mm to 2.50 x 2.50 mm. We compared the MR measurements with the temporally averaged Doppler spectral peak velocities.

Results: For one pixel (r > 0.99: MR flow velocity for pixel dimension 1.00 x 1.00 mm = 1.03x + 9.8 cm/s), the linear correlation was excellent between flow velocities by MR and Doppler guidewire methods. However, for the five adjacent pixels, MR measurements were significantly underestimated using pixels 1.25 x 1.25 mm to 2.50 x 2.50 mm and for mean velocities for all pixel dimensions.

Conclusion: Relatively high spatial resolution allows accurate MR measurement of constant flow velocity in a small tube at spatial peak velocities for one pixel.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Lancet. 1986 Feb 8;1(8476):307-10 - PubMed
    1. Circulation. 1992 Jul;86(1):232-46 - PubMed
    1. Circulation. 2003 Apr 8;107(13):1738-43 - PubMed
    1. J Magn Reson Imaging. 1993 Sep-Oct;3(5):699-703 - PubMed
    1. Scand Cardiovasc J. 2002 May;36(3):180-6 - PubMed

Publication types

MeSH terms

LinkOut - more resources