Exposure rates in high-level-control fluoroscopy for image enhancement
- PMID: 1994395
- DOI: 10.1148/radiology.178.3.1994395
Exposure rates in high-level-control fluoroscopy for image enhancement
Abstract
High-level fluoroscopic boost options that exceed conventional exposure limits are available as a means of reducing quantum mottle during angiography. Federal law does not specify exposure limits for such high-level controls but requires specific means of activation to safeguard against inadvertent use. The American Association of Physicists in Medicine recently recommended that high-level exposure rates not exceed 2.58 mC/kg/min (10R/min). At six institution surveyed, maximum exposure rates ranged from 5.42 to 24 mC/kg/min (21-93 R/min). Activation of high-level capability varied from a simple foot switch to a keyed interlock requiring a second operator to engage. There appears to be no industry coherence in high-level control exposure limits as yet, although the Center for Devices and Radiological Health recently initiated an investigatory program.
Comment in
-
Patient and physician radiation exposure during fluoroscopy.Radiology. 1992 Jan;182(1):286. doi: 10.1148/radiology.182.1.1727302. Radiology. 1992. PMID: 1727302 No abstract available.
-
Absorbed dose in imaging: why measure it?Radiology. 1991 Mar;178(3):622-3. doi: 10.1148/radiology.178.3.1994387. Radiology. 1991. PMID: 1994387 No abstract available.