Comparative healing of surgical incisions created by the PEAK PlasmaBlade, conventional electrosurgery, and a scalpel
- PMID: 19952641
- DOI: 10.1097/PRS.0b013e3181bcee87
Comparative healing of surgical incisions created by the PEAK PlasmaBlade, conventional electrosurgery, and a scalpel
Abstract
Background: The PEAK PlasmaBlade is a new electrosurgical device that uses pulsed radiofrequency to generate a plasma-mediated discharge along the exposed rim of an insulated blade, creating an effective cutting edge while the blade stays near body temperature.
Methods: Full-thickness incisions were made on the dorsums of pigs with the PlasmaBlade, a conventional electrosurgical device, and a scalpel, and blood loss was quantified. Wounds were harvested at designated time points, tested for wound tensile strength, and examined histologically for scar formation and tissue damage.
Results: Bleeding was reduced significantly (59 percent) in PlasmaBlade incisions compared with scalpel incisions, and acute thermal damage from the PlasmaBlade (66 +/- 5 microm) was significantly less than both cut and coagulation mode electrosurgical incisions (456 +/- 35 microm and 615 +/- 22 microm, respectively). Histologic scoring for injury and wound strength was equivalent between the PlasmaBlade and scalpel incisions. By 6 weeks, the healed PlasmaBlade and scalpel incisions were approximately three times stronger, and scar cosmetic appearance was significantly better compared with electrosurgical incisions.
Conclusions: The PlasmaBlade is a promising new surgical instrument that provides atraumatic, scalpel-like cutting precision and electrosurgical-like hemostasis, resulting in minimal bleeding, tissue injury, and scar formation.
Similar articles
-
Comparative healing of human cutaneous surgical incisions created by the PEAK PlasmaBlade, conventional electrosurgery, and a standard scalpel.Plast Reconstr Surg. 2011 Jul;128(1):104-111. doi: 10.1097/PRS.0b013e31821741ed. Plast Reconstr Surg. 2011. PMID: 21701326 Clinical Trial.
-
Comparative study of wound healing in rat skin following incision with a novel picosecond infrared laser (PIRL) and different surgical modalities.Lasers Surg Med. 2016 Apr;48(4):385-91. doi: 10.1002/lsm.22498. Epub 2016 Mar 4. Lasers Surg Med. 2016. PMID: 26941063 Free PMC article.
-
Skeletonized internal thoracic artery harvesting: a low thermal damage electrosurgical device provides improved endothelial layer and tendency to better integrity of the vessel wall compared to conventional electrosurgery.J Cardiothorac Surg. 2018 Oct 11;13(1):105. doi: 10.1186/s13019-018-0797-3. J Cardiothorac Surg. 2018. PMID: 30305183 Free PMC article. Clinical Trial.
-
Pulsed Electron Avalanche Knife (PEAK) PlasmaBlade versus Traditional Electrocautery for Surgery: A Review of Clinical Effectiveness and Cost-Effectiveness [Internet].Ottawa (ON): Canadian Agency for Drugs and Technologies in Health; 2019 Aug 9. Ottawa (ON): Canadian Agency for Drugs and Technologies in Health; 2019 Aug 9. PMID: 31693325 Free Books & Documents. Review.
-
Plasma Scalpels: Devices, Diagnostics, and Applications.Biomedicines. 2022 Nov 17;10(11):2967. doi: 10.3390/biomedicines10112967. Biomedicines. 2022. PMID: 36428535 Free PMC article. Review.
Cited by
-
Mediastinal lymph node evaluation, especially at station 4L, in left upper lobe lung cancer.J Thorac Dis. 2022 Sep;14(9):3321-3334. doi: 10.21037/jtd-22-537. J Thorac Dis. 2022. PMID: 36245624 Free PMC article.
-
Preclinical Comparison of Thermal Tissue Effects from Traditional Electrosurgery and a Low-Temperature Electrosurgical Device during Anterior Cervical Discectomy and Fusion.Int J Spine Surg. 2018 Aug 31;12(4):483-489. doi: 10.14444/5059. eCollection 2018 Aug. Int J Spine Surg. 2018. PMID: 30276109 Free PMC article.
-
Comparative Morphological Effects of Cold-Blade, Electrosurgical, and Plasma Scalpels on Dog Skin.Vet Sci. 2020 Jan 12;7(1):8. doi: 10.3390/vetsci7010008. Vet Sci. 2020. PMID: 31940962 Free PMC article.
-
PlasmaBlade versus Electrocautery for Deep Inferior Epigastric Perforator Flap Harvesting in Autologous Breast Reconstruction: A Comparative Clinical Outcome Study.J Clin Med. 2024 Apr 19;13(8):2388. doi: 10.3390/jcm13082388. J Clin Med. 2024. PMID: 38673669 Free PMC article.
-
Thermally damaged porcine skin is not a surrogate mechanical model of human skin.Sci Rep. 2022 Mar 16;12(1):4565. doi: 10.1038/s41598-022-08551-z. Sci Rep. 2022. PMID: 35296755 Free PMC article.
References
-
- Life Science Intelligence. U.S. Markets for Electrosurgical and Thermal Ablation Products. Irvine, Calif.: Medtech Insight; 2006.
-
- Brown DB. Concepts, considerations, and concerns on the cutting edge of radiofrequency ablation. J Vasc Interv Radiol. 2005;16:597–613.
-
- Massarweh NN, Cosgriff N, Slakey DP. Electrosurgery: History, principles, and current and future uses. J Am Coll Surg. 2006;202:520–530.
-
- Arashiro DS, Rapley JW, Cobb CM, Killoy WJ. Histologic evaluation of porcine skin incisions produced by CO2 laser, electrosurgery, and scalpel. Int J Periodontics Restorative Dent. 1996;16:479–491.
-
- Butler PE, Barry-Walsh C, Curren B, Grace PA, Leader M, Bouchier-Hayes D. Improved wound healing with a modified electrosurgical electrode. Br J Plast Surg. 1991;44:495–499.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical