Perturbation analysis of the effects of B1+ errors on parallel excitation in MRI
- PMID: 19964100
- DOI: 10.1109/IEMBS.2009.5333196
Perturbation analysis of the effects of B1+ errors on parallel excitation in MRI
Abstract
Design of RF pulses for parallel excitation using phased array transmit coils in MRI requires the B1+ maps that are estimated from B1+ mapping experiments. This paper characterizes the effects of B1+ mapping errors on the resulting excitation pattern using a small perturbation analysis based on linearization of the Bloch equation. The accuracy of the proposed perturbation analysis is validated by Bloch equation simulations based on experimental B1+ maps. The perturbation analysis builds a transparent connection between the B1+ mapping errors and the resulting excitation errors, and can be used to study the robustness of the designed RF pulses to B1+ mapping errors. The proposed method may also supply useful metrics for designing RF pulses that are robust to B1+ mapping errors.
Similar articles
-
Design of universal parallel-transmit refocusing kT -point pulses and application to 3D T2 -weighted imaging at 7T.Magn Reson Med. 2018 Jul;80(1):53-65. doi: 10.1002/mrm.27001. Epub 2017 Nov 29. Magn Reson Med. 2018. PMID: 29193250
-
Spatial domain method for the design of RF pulses in multicoil parallel excitation.Magn Reson Med. 2006 Sep;56(3):620-9. doi: 10.1002/mrm.20978. Magn Reson Med. 2006. PMID: 16894579
-
Additive angle method for fast large-tip-angle RF pulse design in parallel excitation.Magn Reson Med. 2008 Apr;59(4):779-87. doi: 10.1002/mrm.21510. Magn Reson Med. 2008. PMID: 18383288
-
Phase-sensitive B1 mapping: Effects of relaxation and RF spoiling.Magn Reson Med. 2018 Jul;80(1):101-111. doi: 10.1002/mrm.27009. Epub 2017 Nov 20. Magn Reson Med. 2018. PMID: 29159883 Free PMC article.
-
Parallel RF transmission in MRI.NMR Biomed. 2006 May;19(3):393-400. doi: 10.1002/nbm.1049. NMR Biomed. 2006. PMID: 16705630 Review.
Cited by
-
Multi-channel microstrip transceiver arrays using harmonics for high field MR imaging in humans.IEEE Trans Med Imaging. 2012 Feb;31(2):183-91. doi: 10.1109/TMI.2011.2166273. Epub 2011 Aug 30. IEEE Trans Med Imaging. 2012. PMID: 21878410 Free PMC article.
-
Quadrature transmit array design using single-feed circularly polarized patch antenna for parallel transmission in MR imaging.Quant Imaging Med Surg. 2014 Feb;4(1):11-8. doi: 10.3978/j.issn.2223-4292.2014.02.03. Quant Imaging Med Surg. 2014. PMID: 24649430 Free PMC article.
-
Robust and fast whole brain mapping of the RF transmit field B1 at 7T.PLoS One. 2012;7(3):e32379. doi: 10.1371/journal.pone.0032379. Epub 2012 Mar 12. PLoS One. 2012. PMID: 22427831 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Medical