A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion
- PMID: 1996749
- DOI: 10.1111/j.1365-2044.1991.tb09303.x
A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion
Abstract
The performance of 20 pulse oximeters with finger probes was evaluated by comparison of their readings with directly measured arterial blood oxygen saturations. The samples were taken from patients who had undergone cardiac surgery under hypothermic cardiopulmonary bypass and had poor peripheral perfusion. The mean difference (bias, accuracy), standard deviation (precision) and drop-out rate for each pulse oximeter was determined. An overall ranking of performance of each pulse oximeter was calculated using five criteria (accuracy, precision, number of readings within 3% of standard, percentage of readings given within 3% of standard, expected overread limit in 95% of cases). Two pulse oximeters achieved a combination of accuracy and precision such that 95% of measurements would be expected to be within 4% of the co-oximeter value; these two also had the lowest drop-out rate.
Comment in
-
'Finger-painting' and the pulse oximeter.Anaesthesia. 1991 May;46(5):420-1. doi: 10.1111/j.1365-2044.1991.tb09572.x. Anaesthesia. 1991. PMID: 1867708 No abstract available.
-
Pulse oximeters and poor perfusion.Anaesthesia. 1991 Oct;46(10):887. doi: 10.1111/j.1365-2044.1991.tb09614.x. Anaesthesia. 1991. PMID: 1952011 No abstract available.
-
Pulse oximetry in the postcardiopulmonary bypass period.Anaesthesia. 1991 Jun;46(6):517-8. doi: 10.1111/j.1365-2044.1991.tb11731.x. Anaesthesia. 1991. PMID: 2048696 No abstract available.
Similar articles
-
Pulse oximeter probes. A comparison between finger, nose, ear and forehead probes under conditions of poor perfusion.Anaesthesia. 1991 Apr;46(4):260-5. doi: 10.1111/j.1365-2044.1991.tb11492.x. Anaesthesia. 1991. PMID: 2024741
-
Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia.Anesth Analg. 2010 Jul;111(1):143-8. doi: 10.1213/ANE.0b013e3181c91bb6. Epub 2009 Dec 10. Anesth Analg. 2010. PMID: 20007731
-
Continuous intra-arterial oximetry, pulse oximetry, and co-oximetry during cardiac surgery.J Cardiothorac Vasc Anesth. 1992 Dec;6(6):668-73. doi: 10.1016/1053-0770(92)90048-c. J Cardiothorac Vasc Anesth. 1992. PMID: 1472661
-
Potential errors in pulse oximetry. II. Effects of changes in saturation and signal quality.Anaesthesia. 1991 Mar;46(3):207-12. doi: 10.1111/j.1365-2044.1991.tb09411.x. Anaesthesia. 1991. PMID: 2014899 Review.
-
Potential errors in pulse oximetry. I. Pulse oximeter evaluation.Anaesthesia. 1991 Mar;46(3):202-6. doi: 10.1111/j.1365-2044.1991.tb09410.x. Anaesthesia. 1991. PMID: 2014898 Review.
Cited by
-
A system for investigating oesophageal photoplethysmographic signals in anaesthetised patients.Med Biol Eng Comput. 1999 Sep;37(5):639-43. doi: 10.1007/BF02513360. Med Biol Eng Comput. 1999. PMID: 10723903 Clinical Trial.
-
Pulse oximetry: a practical review.BMJ. 1995 Aug 5;311(7001):367-70. doi: 10.1136/bmj.311.7001.367. BMJ. 1995. PMID: 7640545 Free PMC article. Review. No abstract available.
-
Failure rate of pulse oximetry in the postanesthesia care unit.J Clin Monit. 1993 Nov;9(5):326-9. doi: 10.1007/BF01618673. J Clin Monit. 1993. PMID: 8106885
-
Racial effects on masimo pulse oximetry: impact of low perfusion index.J Clin Monit Comput. 2024 Apr;38(2):347-354. doi: 10.1007/s10877-023-01113-2. Epub 2024 Jan 19. J Clin Monit Comput. 2024. PMID: 38238634 Free PMC article.
-
Limitations of SpO2 / FiO2-ratio for classification and monitoring of acute respiratory distress syndrome-an observational cohort study.Crit Care. 2025 Feb 19;29(1):82. doi: 10.1186/s13054-025-05317-7. Crit Care. 2025. PMID: 39972458 Free PMC article.