Glistenings in a large series of hydrophobic acrylic intraocular lenses
- PMID: 19969218
- DOI: 10.1016/j.jcrs.2009.06.029
Glistenings in a large series of hydrophobic acrylic intraocular lenses
Abstract
Purpose: To evaluate the incidence and severity of glistenings in hydrophobic acrylic intraocular lenses (IOLs) and assess the potential correlation between glistenings and clinical and demographic factors.
Setting: Service d'Ophtalmologie, Centre Hospitalier et Universitaire Bordeaux, Bordeaux, France.
Methods: Eligible patients received an AcrySof IOL between September 2000 and December 2007 and had a routine ophthalmic visit between January 2007 and March 2008. The incidence and severity of glistenings were graded subjectively and analyzed in relation to patient age and sex, length of follow-up, IOL model, IOL power, neodymium:YAG laser capsulotomy, corrected distance visual acuity (CDVA), spherical equivalent (SE), and selected ocular and systemic diseases and medications.
Results: Glistenings occurred in 157 (60.4%) of the 260 eyes; they were absent in 103 eyes (39.6%). Glistenings were of grade 1 severity in 87 eyes (33.5%) and of grade 2 severity in 70 eyes (26.9%).
Conclusions: The results suggest a potential association between the incidence of glistenings and IOL power and glaucoma, but not between glistenings and age, sex, IOL model, length of follow-up, CDVA, SE, or most ocular and systemic diseases and medications.