Time-resolved in vivo luminescence dosimetry for online error detection in pulsed dose-rate brachytherapy
- PMID: 19994514
- DOI: 10.1118/1.3238102
Time-resolved in vivo luminescence dosimetry for online error detection in pulsed dose-rate brachytherapy
Abstract
Purpose: The purpose of this study is to present and evaluate a dose-verification protocol for pulsed dose-rate (PDR) brachytherapy based on in vivo time-resolved (1 s time resolution) fiber-coupled luminescence dosimetry.
Methods: Five cervix cancer patients undergoing PDR brachytherapy (Varian GammaMed Plus with 192Ir) were monitored. The treatments comprised from 10 to 50 pulses (1 pulse/h) delivered by intracavitary/interstitial applicators (tandem-ring systems and/or needles). For each patient, one or two dosimetry probes were placed directly in or close to the tumor region using stainless steel or titanium needles. Each dosimeter probe consisted of a small aluminum oxide crystal attached to an optical fiber cable (1 mm outer diameter) that could guide radioluminescence (RL) and optically stimulated luminescence (OSL) from the crystal to special readout instrumentation. Positioning uncertainty and hypothetical dose-delivery errors (interchanged guide tubes or applicator movements from +/-5 to +/-15 mm) were simulated in software in order to assess the ability of the system to detect errors.
Results: For three of the patients, the authors found no significant differences (P>0.01) for comparisons between in vivo measurements and calculated reference values at the level of dose per dwell position, dose per applicator, or total dose per pulse. The standard deviations of the dose per pulse were less than 3%, indicating a stable dose delivery and a highly stable geometry of applicators and dosimeter probes during the treatments. For the two other patients, the authors noted significant deviations for three individual pulses and for one dosimeter probe. These deviations could have been due to applicator movement during the treatment and one incorrectly positioned dosimeter probe, respectively. Computer simulations showed that the likelihood of detecting a pair of interchanged guide tubes increased by a factor of 10 or more for the considered patients when going from integrating to time-resolved dose verification. The likelihood of detecting a +/-15 mm displacement error increased by a factor of 1.5 or more.
Conclusions: In vivo fiber-coupled RL/OSL dosimetry based on detectors placed in standard brachytherapy needles was demonstrated. The time-resolved dose-rate measurements were found to provide a good way to visualize the progression and stability of PDR brachytherapy dose delivery, and time-resolved dose-rate measurements provided an increased sensitivity for detection of dose-delivery errors compared with time-integrated dosimetry.
Similar articles
-
Characterization of a fiber-coupled Al2O3:C luminescence dosimetry system for online in vivo dose verification during 192Ir brachytherapy.Med Phys. 2009 Mar;36(3):708-18. doi: 10.1118/1.3063006. Med Phys. 2009. PMID: 19378731
-
Identifying afterloading PDR and HDR brachytherapy errors using real-time fiber-coupled Al(2)O(3):C dosimetry and a novel statistical error decision criterion.Radiother Oncol. 2011 Sep;100(3):456-62. doi: 10.1016/j.radonc.2011.09.009. Epub 2011 Sep 29. Radiother Oncol. 2011. PMID: 21963285
-
Evaluation of a real-time BeO ceramic fiber-coupled luminescence dosimetry system for dose verification of high dose rate brachytherapy.Med Phys. 2015 Nov;42(11):6349-56. doi: 10.1118/1.4931968. Med Phys. 2015. PMID: 26520726
-
In vivo dosimetry in brachytherapy.Med Phys. 2013 Jul;40(7):070902. doi: 10.1118/1.4810943. Med Phys. 2013. PMID: 23822403 Review.
-
In vivo dosimetry in external beam radiotherapy.Med Phys. 2013 Jul;40(7):070903. doi: 10.1118/1.4811216. Med Phys. 2013. PMID: 23822404 Review.
Cited by
-
The clinical application of in vivo dosimetry for gynaecological brachytherapy: A scoping review.Tech Innov Patient Support Radiat Oncol. 2024 Nov 16;33:100290. doi: 10.1016/j.tipsro.2024.100290. eCollection 2025 Mar. Tech Innov Patient Support Radiat Oncol. 2024. PMID: 39802319 Free PMC article. Review.
-
On the nature of the light produced within PMMA optical light guides in scintillation fiber-optic dosimetry.Phys Med Biol. 2013 Apr 7;58(7):2073-84. doi: 10.1088/0031-9155/58/7/2073. Epub 2013 Mar 8. Phys Med Biol. 2013. PMID: 23470253 Free PMC article.
-
Time-resolved dose distributions to moving targets during volumetric modulated arc therapy with and without dynamic MLC tracking.Med Phys. 2013 Nov;40(11):111723. doi: 10.1118/1.4826161. Med Phys. 2013. PMID: 24320431 Free PMC article.
-
In vivo dosimetry: trends and prospects for brachytherapy.Br J Radiol. 2014 Sep;87(1041):20140206. doi: 10.1259/bjr.20140206. Epub 2014 Jul 8. Br J Radiol. 2014. PMID: 25007037 Free PMC article. Review.
-
A phantom study of an in vivo dosimetry system using plastic scintillation detectors for real-time verification of 192Ir HDR brachytherapy.Med Phys. 2011 May;38(5):2542-51. doi: 10.1118/1.3572229. Med Phys. 2011. PMID: 21776789 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources