Treatment planning for volumetric modulated arc therapy
- PMID: 19994523
- DOI: 10.1118/1.3240488
Treatment planning for volumetric modulated arc therapy
Abstract
Purpose: Volumetric modulated arc therapy (VMAT) is a specific type of intensity-modulated radiation therapy (IMRT) in which the gantry speed, multileaf collimator (MLC) leaf position, and dose rate vary continuously during delivery. A treatment planning system for VMAT is presented.
Methods: Arc control points are created uniformly throughout one or more arcs. An iterative least-squares algorithm is used to generate a fluence profile at every control point. The control points are then grouped and all of the control points in a given group are used to approximate the fluence profiles. A direct-aperture optimization is then used to improve the solution, taking into account the allowed range of leaf motion of the MLC. Dose is calculated using a fast convolution algorithm and the motion between control points is approximated by 100 interpolated dose calculation points. The method has been applied to five cases, consisting of lung, rectum, prostate and seminal vesicles, prostate and pelvic lymph nodes, and head and neck. The resulting plans have been compared with segmental (step-and-shoot) IMRT and delivered and verified on an Elekta Synergy to ensure practicality.
Results: For the lung, prostate and seminal vesicles, and rectum cases, VMAT provides a plan of similar quality to segmental IMRT but with faster delivery by up to a factor of 4. For the prostate and pelvic nodes and head-and-neck cases, the critical structure doses are reduced with VMAT, both of these cases having a longer delivery time than IMRT. The plans in general verify successfully, although the agreement between planned and measured doses is not very close for the more complex cases, particularly the head-and-neck case.
Conclusions: Depending upon the emphasis in the treatment planning, VMAT provides treatment plans which are higher in quality and/or faster to deliver than IMRT. The scheme described has been successfully introduced into clinical use.
Similar articles
-
A comprehensive formulation for volumetric modulated arc therapy planning.Med Phys. 2016 Jul;43(7):4263. doi: 10.1118/1.4953832. Med Phys. 2016. PMID: 27370141 Free PMC article.
-
Analysis of direct clinical consequences of MLC positional errors in volumetric-modulated arc therapy using 3D dosimetry system.J Appl Clin Med Phys. 2015 Sep 8;16(5):296–305. doi: 10.1120/jacmp.v16i5.5515. J Appl Clin Med Phys. 2015. PMID: 26699311 Free PMC article.
-
Impact of leaf motion constraints on IMAT plan quality, deliver accuracy, and efficiency.Med Phys. 2011 Nov;38(11):6106-18. doi: 10.1118/1.3651698. Med Phys. 2011. PMID: 22047375
-
Utilising the Virtual Environment for Radiotherapy Training System to Support Undergraduate Teaching of IMRT, VMAT, DCAT Treatment Planning, and QA Concepts.J Med Imaging Radiat Sci. 2018 Mar;49(1):31-38. doi: 10.1016/j.jmir.2017.11.002. Epub 2018 Jan 3. J Med Imaging Radiat Sci. 2018. PMID: 30479285 Review.
-
Potential advantages of volumetric arc therapy in head and neck cancer.Head Neck. 2015 Jun;37(6):909-14. doi: 10.1002/hed.23685. Epub 2014 Apr 30. Head Neck. 2015. PMID: 24623665 Review.
Cited by
-
Split X-jaw Planning Technique of Volumetric Modulated Arc Radiotherapy.J Med Phys. 2023 Oct-Dec;48(4):398-401. doi: 10.4103/jmp.jmp_67_23. Epub 2023 Dec 5. J Med Phys. 2023. PMID: 38223790 Free PMC article.
-
Noninvasive treatment of two deeply invasive cutaneous squamous cell carcinomas of the midface located close to the orbits with intensity-modulated arc therapy: A case report.Clin Case Rep. 2022 Jul 25;10(7):e6149. doi: 10.1002/ccr3.6149. eCollection 2022 Jul. Clin Case Rep. 2022. PMID: 35898739 Free PMC article.
-
Dual-Modality X-Ray-Induced Radiation Acoustic and Ultrasound Imaging for Real-Time Monitoring of Radiotherapy.BME Front. 2020 May 26;2020:9853609. doi: 10.34133/2020/9853609. eCollection 2020. BME Front. 2020. PMID: 37849969 Free PMC article.
-
Dosimetric comparison between intensity-modulated radiation therapy and volumetric-modulated arc therapy to enhance bladder and bowel.J Med Life. 2023 Sep;16(9):1381-1387. doi: 10.25122/jml-2022-0317. J Med Life. 2023. PMID: 38107703 Free PMC article.
-
A convex optimization approach to radiation treatment planning with dose constraints.Optim Eng. 2019 Mar;20(1):277-300. doi: 10.1007/s11081-018-9409-2. Epub 2018 Nov 22. Optim Eng. 2019. PMID: 37990749 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources