Development of protein A functionalized microcantilever immunosensors for the analyses of small molecules at parts per trillion levels
- PMID: 19994853
- DOI: 10.1021/ac901937g
Development of protein A functionalized microcantilever immunosensors for the analyses of small molecules at parts per trillion levels
Abstract
Development of microcantilever biosensors for small molecules was exemplified with the beta-adrenergic agonist clenbuterol and the antibiotic chloramphenicol. In this paper, antibody sulfhydrylation and protein A were used to modify the microcantilever Au surface, and the antibody activities on the microcantilever were evaluated with direct competitive enzyme-linked immunosorbent assay (dcELISA). The activity of the antibodies immobilized on the microcantilever via protein A was 1.7-fold of that via the sulfhydrylation reagent 2-iminothiolane hydrochloride. A microcantilever immunosensor method with protein A as the functionalization reagent was established to detect the residues of clenbuterol and chloramphenicol at limits of detection (LOD) of approximately 0.1 and 0.2 ng/mL, respectively. Such LODs were better than that of the corresponding dcELISAs. The concentration of clenbuterol in a fortified feed sample detected with the microcantilever immunosensor after thorough extraction and purification agreed well with that detected with the dcELISA. Protein A showed to be simple and reproducible for functionalization of the antibodies on the Au surface and, thus, has common application values in microcantilever immunosensor development. The results suggest that microcantilever immunosensors be suitable for detection of small molecules, and the assay sensitivity is mainly related to the quality and activities of the antibodies.
Similar articles
-
Small molecule microarrays for drug residue detection in foodstuffs.J Agric Food Chem. 2006 Sep 20;54(19):6978-83. doi: 10.1021/jf061105+. J Agric Food Chem. 2006. PMID: 16968051
-
Simultaneous detection for three kinds of veterinary drugs: chloramphenicol, clenbuterol and 17-beta-estradiol by high-throughput suspension array technology.Anal Chim Acta. 2009 Jan 19;632(1):128-34. doi: 10.1016/j.aca.2008.10.061. Epub 2008 Nov 5. Anal Chim Acta. 2009. PMID: 19100892
-
Improvement of antibody immobilization using hyperbranched polymer and protein A.Anal Biochem. 2011 Feb 1;409(1):22-7. doi: 10.1016/j.ab.2010.09.028. Epub 2010 Sep 24. Anal Biochem. 2011. PMID: 20869942
-
Electrochemical immunochip sensor for aflatoxin M1 detection.Anal Chem. 2009 Jul 1;81(13):5291-8. doi: 10.1021/ac900511e. Anal Chem. 2009. PMID: 19489595
-
A Genosensor Based on the Modification of a Microcantilever: A Review.Micromachines (Basel). 2023 Feb 10;14(2):427. doi: 10.3390/mi14020427. Micromachines (Basel). 2023. PMID: 36838127 Free PMC article. Review.
Cited by
-
Highly sensitive nanomechanical assay for the stress transmission of carbon chain.Sens Actuators B Chem. 2013 Sep;186:353-359. doi: 10.1016/j.snb.2013.06.035. Epub 2013 Jun 19. Sens Actuators B Chem. 2013. PMID: 32288244 Free PMC article.
-
Recent developments in emerging microimmunoassays.Bioanalysis. 2013 Jan;5(2):245-64. doi: 10.4155/bio.12.298. Bioanalysis. 2013. PMID: 23330564 Free PMC article. Review.
-
Molecular Recognition of Surface Trans-Sialidases in Extracellular Vesicles of the Parasite Trypanosoma cruzi Using Atomic Force Microscopy (AFM).Int J Mol Sci. 2022 Jun 28;23(13):7193. doi: 10.3390/ijms23137193. Int J Mol Sci. 2022. PMID: 35806197 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources