Skin reflectance pulse oximetry: in vivo measurements from the forearm and calf
- PMID: 1999702
- DOI: 10.1007/BF01617892
Skin reflectance pulse oximetry: in vivo measurements from the forearm and calf
Abstract
This study describes the results from a series of human experiments demonstrating the ability to measure arterial hemoglobin oxygen saturation (SaO2) from the forearm and calf using a reflectance pulse oximeter sensor. A special optical reflectance sensor that includes a heating element was interfaced to a temperature controller and a commercial Data-scope ACCUSAT pulse oximeter that was adapted for this study to perform as a reflectance pulse oximeter. The reflectance pulse oximeter sensor was evaluated in a group of 10 healthy adult volunteers during steady-state hypoxia. Hypoxia was induced by gradually lowering the inspired fraction of oxygen in the breathing gas mixture from 100 to 12%. Simultaneous SaO2 measurements obtained from the forearm and calf with two identical reflectance pulse oximeters were compared with SaO2 values measured by a finger sensor that was interfaced to a standard Datascope ACCUSAT transmittance pulse oximeter. The equations for the best-fitted linear regression lines between the percent reflectance, SpO2(r), and transmittance, SpO2(t), values in the range between 73 and 100% were SpO2(r) = -7.06 + 1.09 SpO2(t) for the forearm (n = 91, r = 0.95) and SpO2(r) = 7.78 + 0.93 SpO2(t) for the calf (n = 93, r = 0.88). The regression analysis of the forearm data revealed a mean +/- SD error of 2.47 +/- 1.66% (SaO2 = 90-100%), 2.35 +/- 2.45% (SaO2 = 80-89%), and 2.42 +/- 1.20% (SaO2 = 70-79%). The corresponding regression analysis of the calf data revealed a mean +/- SD error of 3.36 +/- 3.06% (SaO2 = 90-100%), 3.45 +/- 4.12% (SaO2 = 80-89%), and 2.97 +/- 2.75% (SaO2 = 70-79%).(ABSTRACT TRUNCATED AT 250 WORDS)
Similar articles
-
Experimental and clinical evaluation of a noninvasive reflectance pulse oximeter sensor.J Clin Monit. 1992 Oct;8(4):257-66. doi: 10.1007/BF01617907. J Clin Monit. 1992. PMID: 1453185
-
Noninvasive measurement of arterial oxyhemoglobin saturation with a heated and a non-heated skin reflectance pulse oximeter sensor.Biomed Instrum Technol. 1991 Nov-Dec;25(6):472-80. Biomed Instrum Technol. 1991. PMID: 1777770
-
Design and evaluation of a new reflectance pulse oximeter sensor.Med Instrum. 1988 Aug;22(4):167-73. Med Instrum. 1988. PMID: 3173170
-
Impacts of Skin Color and Hypoxemia on Noninvasive Assessment of Peripheral Blood Oxygen Saturation: A Scoping Review.Cureus. 2023 Sep 27;15(9):e46078. doi: 10.7759/cureus.46078. eCollection 2023 Sep. Cureus. 2023. PMID: 37900526 Free PMC article.
-
Pulse oximetry: fundamentals and technology update.Med Devices (Auckl). 2014 Jul 8;7:231-9. doi: 10.2147/MDER.S47319. eCollection 2014. Med Devices (Auckl). 2014. PMID: 25031547 Free PMC article. Review.
Cited by
-
Experimental and clinical evaluation of a noninvasive reflectance pulse oximeter sensor.J Clin Monit. 1992 Oct;8(4):257-66. doi: 10.1007/BF01617907. J Clin Monit. 1992. PMID: 1453185
-
Reflectance pulse oximetry at the forehead of newborns: the influence of varying pressure on the probe.J Clin Monit. 1996 Nov;12(6):421-8. doi: 10.1007/BF02199702. J Clin Monit. 1996. PMID: 8982906
-
The effect of pulsating arteries on reflectance pulse oximetry: measurements in adults and neonates.J Clin Monit. 1995 Mar;11(2):118-22. doi: 10.1007/BF01617733. J Clin Monit. 1995. PMID: 7760084
-
Reflectance pulse oximetry at the forehead improves by pressure on the probe.J Clin Monit. 1995 Jul;11(4):237-44. doi: 10.1007/BF01617518. J Clin Monit. 1995. PMID: 7561997
References
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Other Literature Sources