Comparison of anterior chamber depth measurements using Galilei, HR Pentacam, and Orbscan II
- PMID: 20004876
- DOI: 10.1016/j.optm.2009.04.100
Comparison of anterior chamber depth measurements using Galilei, HR Pentacam, and Orbscan II
Abstract
Background: Anterior chamber depth measurement is used frequently in cataract and refractive surgery for advanced intraocular lens (IOL) power calculation formulas and implantation of phakic IOLs. In this article, we show agreement in anterior chamber depth measurements in normal eyes among 3 noncontact systems, namely Galilei, HR Pentacam, and Orbscan II.
Methods: The anterior chamber depth measurements were obtained with Galilei, HR Pentacam, and the Orbscan II in both eyes of 37 healthy subjects. Pearson's correlations were calculated, and mean anterior chamber depth measurements were compared. Bland-Altman plots were used to assess the difference between individual measurements for each patient.
Results: The mean anterior chamber depth readings (+/- standard error of means) with Galilei, Pentacam, and Orbscan were 3.22 +/- 0.05, 3.25 +/- 0.05, and 3.54 +/- 0.07, respectively. The difference between Galilei and Pentacam was not statistically significant (P = 0.013). However, Orbscan measurements were larger than those of either Galilei or Pentacam (P < 0.001). The differences between Orbscan and either Galilei or Pentacam were increased with greater anterior chamber depths. The 95% limits of agreement between Orbscan and Galilei, Orbscan and Pentacam, and Pentacam and Galilei in different ranges of chamber depth were ([ < 3 mm, -0.43 to +0.71 mm]; [3 to 3.5 mm, -0.23 to +0.83]; [ > 3.5 mm, +0.41 to +0.53]), ([ < 3 mm, -0.43 to +0.71 mm]; [3 to 3.5 mm, -0.26 to +0.80]; [ > 3.5 mm, +0.38 to +0.54]), and ([ < 3 mm, -0.05 to +0.06 mm]; [3 to 3.5 mm, -0.06 to +0.10]; [ > 3.5 mm, -0.11 to +0.12]), respectively.
Conclusions: These data indicate that in the assessment of normal eyes, Orbscan gives consistently higher measurements for anterior chamber depth compared with Galilei and Pentacam. Because the differences between Orbscan and either Galilei or Pentacam were not within clinically acceptable levels, they are not interchangeable in every clinical situation. However, the differences among anterior chamber depth values measured with Galilei and Pentacam were within clinically acceptable levels. Thus, these 2 systems can be regarded as interchangeable.
Similar articles
-
Comparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyes.Acta Ophthalmol. 2009 Jun;87(4):386-91. doi: 10.1111/j.1755-3768.2008.01278.x. Epub 2008 Sep 5. Acta Ophthalmol. 2009. PMID: 18778337
-
Anterior chamber depth measurement in phakic and pseudophakic eyes.Optom Vis Sci. 2008 Dec;85(12):1193-200. doi: 10.1097/OPX.0b013e31818e8ceb. Optom Vis Sci. 2008. PMID: 19050474
-
Central corneal thickness, anterior chamber depth, and pupil diameter measurements using Visante OCT, Orbscan, and Pentacam.J Refract Surg. 2010 Feb;26(2):127-33. doi: 10.3928/1081597X-20100121-08. Epub 2010 Feb 12. J Refract Surg. 2010. PMID: 20163077
-
[Pentacam--the new way for anterior eye segment imaging and mapping].Klin Oczna. 2005;107(10-12):728-31. Klin Oczna. 2005. PMID: 16619832 Review. Polish.
-
Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.Int J Ophthalmol. 2016 Jul 18;9(7):1057-65. doi: 10.18240/ijo.2016.07.20. eCollection 2016. Int J Ophthalmol. 2016. PMID: 27500117 Free PMC article. Review.
Cited by
-
Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.Int J Ophthalmol. 2015 Apr 18;8(2):399-402. doi: 10.3980/j.issn.2222-3959.2015.02.33. eCollection 2015. Int J Ophthalmol. 2015. PMID: 25938063 Free PMC article.
-
Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster.J Ophthalmol. 2012;2012:430249. doi: 10.1155/2012/430249. Epub 2012 Jan 19. J Ophthalmol. 2012. PMID: 22315660 Free PMC article.
-
Performance of Zernike polynomials in reconstructing raw-elevation data captured by Pentacam HR, Medmont E300 and Eye Surface Profiler.Heliyon. 2021 Dec 18;7(12):e08623. doi: 10.1016/j.heliyon.2021.e08623. eCollection 2021 Dec. Heliyon. 2021. PMID: 35005275 Free PMC article.
-
Anterior chamber indices in a population-based study using the Pentacam.Int Ophthalmol. 2019 Sep;39(9):2033-2040. doi: 10.1007/s10792-018-1037-5. Epub 2018 Oct 31. Int Ophthalmol. 2019. PMID: 30382464
-
Optimization of biometry for best refractive outcome in cataract surgery.Indian J Ophthalmol. 2024 Jan 1;72(1):29-43. doi: 10.4103/IJO.IJO_1219_23. Epub 2023 Dec 22. Indian J Ophthalmol. 2024. PMID: 38131567 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Medical