Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia
- PMID: 20007731
- DOI: 10.1213/ANE.0b013e3181c91bb6
Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia
Abstract
Background: Methemoglobin in the blood cannot be detected by conventional pulse oximetry, although it can bias the oximeter's estimate (Spo2) of the true arterial functional oxygen saturation (Sao2). A recently introduced "Pulse CO-Oximeter" (Masimo Rainbow SET(R) Radical-7 Pulse CO-Oximeter, Masimo Corp., Irvine, CA) is intended to additionally monitor noninvasively the fractional carboxyhemoglobin and methemoglobin content in blood. The purpose of our study was to determine whether hypoxia affects the new device's estimated methemoglobin reading accuracy, and whether the presence of methemoglobin impairs the ability of the Radical-7 and a conventional pulse oximeter (Nonin 9700, Nonin Medical Inc., Plymouth, MN) to detect decreases in Sao2.
Methods: Eight and 6 healthy adults were included in 2 study groups, respectively, each fitted with multiple sensors and a radial arterial catheter for blood sampling. In the first group, IV administration of approximately 300 mg sodium nitrite increased subjects' methemoglobin level to a 7% to 8% target and hypoxia was induced to different levels of Sao2 (70%-100%) by varying fractional inspired oxygen. In the second group, 15% methemoglobin at room air and 80% Sao2 were targeted. Pulse CO-oximeter readings were compared with arterial blood values measured using a Radiometer multiwavelength hemoximeter. Pulse CO-oximeter methemoglobin reading performance was analyzed by observing the incidence of meaningful reading errors at the various hypoxia levels. This was used to determine the impact on predictive values for detecting methemoglobinemia. Spo2 reading bias, precision, and root mean square error were evaluated during conditions of elevated methemoglobin.
Results: Observations spanned 66.2% to 99% Sao2 and 0.6% to 14.4% methemoglobin over the 2 groups (170 blood draws). Masimo methemoglobin reading bias and precision over the full Sao2 span was 7.7% +/- 13.0%. Best accuracy was found in the 95% to 100% Sao2 range (1.9% +/- 2.5%), progressing to its worst in the 70% to 80% range (24.8% +/- 15.6%). Occurrence of methemoglobin readings in error >5% increased over each 5-point decrease in Sao2 (P < 0.05). Masimo Spo2 readings were biased -6.3% +/- 3.0% in the 95% to 100% Sao2 range with 4% to 8.3% methemoglobin. Both the Radical-7 and Nonin 9700 pulse oximeters accurately detected decreases in Sao(2) <90% with 4% to 15% methemoglobin, despite displaying low Spo2 readings when Sao2 was >95%.
Conclusions: The Radical-7's methemoglobin readings become progressively more inaccurate as Sao2 decreases <95%, at times overestimating true values by 10% to 40%. Elevated methemoglobin causes the Spo2 readings to underestimate Sao2 similar to conventional 2-wavelength pulse oximeters at high saturation. Spo2 readings from both types of instruments continue to trend downward during the development of hypoxemia (Sao2 <90%) with methemoglobin levels up to 15%.
Comment in
-
New sensors should improve Masimo SpMet error found by Feiner et al.Anesth Analg. 2010 Jul;111(1):241; author reply 241. doi: 10.1213/ANE.0b013e3181e2936c. Anesth Analg. 2010. PMID: 20576969 No abstract available.
Similar articles
-
Improved accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia.Anesth Analg. 2010 Nov;111(5):1160-7. doi: 10.1213/ANE.0b013e3181f46da8. Epub 2010 Sep 14. Anesth Analg. 2010. PMID: 20841412
-
Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study.Anesthesiology. 2006 Nov;105(5):892-7. doi: 10.1097/00000542-200611000-00008. Anesthesiology. 2006. PMID: 17065881
-
Accuracy of pulse oximeters in detecting hypoxemia in patients with chronic thromboembolic pulmonary hypertension.PLoS One. 2015 May 15;10(5):e0126979. doi: 10.1371/journal.pone.0126979. eCollection 2015. PLoS One. 2015. PMID: 25978517 Free PMC article.
-
The measurement of dyshemoglobins and total hemoglobin by pulse oximetry.Curr Opin Anaesthesiol. 2008 Dec;21(6):805-10. doi: 10.1097/ACO.0b013e328316bb6f. Curr Opin Anaesthesiol. 2008. PMID: 18997533 Review.
-
The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration.Anesth Analg. 2012 May;114(5):972-8. doi: 10.1213/ANE.0b013e318233041a. Epub 2011 Sep 29. Anesth Analg. 2012. PMID: 21965370 Review.
Cited by
-
Methemoglobinemia and arterial blood gas.Indian Heart J. 2015 Jan-Feb;67(1):63. doi: 10.1016/j.ihj.2015.02.001. Epub 2015 Feb 25. Indian Heart J. 2015. PMID: 25820055 Free PMC article. No abstract available.
-
Case files of the medical toxicology fellowship at Banner Good Samaritan Medical Center in Phoenix, AZ: methemoglobinemia following dapsone exposure.J Med Toxicol. 2011 Jun;7(2):139-46. doi: 10.1007/s13181-011-0151-9. J Med Toxicol. 2011. PMID: 21487934 Free PMC article.
-
Validity of pulse oximetry measures for heart rate and oxygen saturation during profound hypoxia in normobaric simulated extreme altitudes.PLoS One. 2025 Jun 23;20(6):e0326674. doi: 10.1371/journal.pone.0326674. eCollection 2025. PLoS One. 2025. PMID: 40549783 Free PMC article.
-
Evaluation of multiwave pulse total-hemoglobinometer during general anesthesia.J Anesth. 2014 Jun;28(3):463-6. doi: 10.1007/s00540-013-1730-5. Epub 2013 Oct 22. J Anesth. 2014. PMID: 24146037
-
Unusual cause of persistent dyspnea in a patient with nephrotic syndrome: dapsone-induced methemoglobinemia.CEN Case Rep. 2021 Aug;10(3):336-340. doi: 10.1007/s13730-020-00565-8. Epub 2021 Jan 8. CEN Case Rep. 2021. PMID: 33417185 Free PMC article.