Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction
- PMID: 20008999
- DOI: 10.2967/jnumed.109.065532
Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction
Abstract
We aimed at establishing the optimal scan time for nuclear myocardial perfusion imaging (MPI) on an ultrafast cardiac gamma-camera using a novel cadmium-zinc-telluride (CZT) solid-state detector technology.
Methods: Twenty patients (17 male; BMI range, 21.7-35.5 kg/m(2)) underwent 1-d (99m)Tc-tetrofosmin adenosine stress and rest MPI protocols, each with a 15-min acquisition on a standard dual-detector SPECT camera. All scans were immediately repeated on an ultrafast CZT camera over a 6-min acquisition time and reconstructed from list-mode raw data to obtain scan durations of 1 min, 2 min, etc., up to a maximum of 6 min. For each of the scan durations, the segmental tracer uptake value (percentage of maximum myocardial uptake) from the CZT camera was compared by intraclass correlation with standard SPECT camera data using a 20-segment model, and clinical agreement was assessed per coronary territory. Scan durations above which no further relevant improvement in uptake correlation was found were defined as minimal required scan times, for which Bland-Altman limits of agreement were calculated.
Results: Minimal required scan times were 3 min for low dose (r = 0.81; P < 0.001; Bland-Altman, -11.4% to 12.2%) and 2 min for high dose (r = 0.80; P < 0.001; Bland-Altman, -7.6% to 12.9%), yielding a clinical agreement of 95% and 97%, respectively.
Conclusion: We have established the minimal scan time for a CZT solid-state detector system, which allows 1-d stress/rest MPI with a substantially reduced acquisition time resulting in excellent agreement with regard to uptake and clinical findings, compared with MPI from a standard dual-head SPECT gamma-camera.
Similar articles
-
Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation.Eur J Nucl Med Mol Imaging. 2010 Apr;37(4):773-8. doi: 10.1007/s00259-009-1375-7. Epub 2010 Jan 27. Eur J Nucl Med Mol Imaging. 2010. PMID: 20107783
-
Semiconductor detectors allow low-dose-low-dose 1-day SPECT myocardial perfusion imaging.J Nucl Med. 2011 Aug;52(8):1204-9. doi: 10.2967/jnumed.110.085415. J Nucl Med. 2011. PMID: 21810589
-
Validation of CT attenuation correction for high-speed myocardial perfusion imaging using a novel cadmium-zinc-telluride detector technique.J Nucl Med. 2010 Oct;51(10):1539-44. doi: 10.2967/jnumed.110.078170. Epub 2010 Sep 16. J Nucl Med. 2010. PMID: 20847159
-
Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM).Eur J Nucl Med Mol Imaging. 2016 Dec;43(13):2423-2432. doi: 10.1007/s00259-016-3467-5. Epub 2016 Aug 19. Eur J Nucl Med Mol Imaging. 2016. PMID: 27542010 Review.
-
New solid state cadmium-zinc-telluride technology for cardiac single photon emission computed tomographic myocardial perfusion imaging.Expert Rev Med Devices. 2017 Mar;14(3):213-222. doi: 10.1080/17434440.2017.1296763. Epub 2017 Mar 1. Expert Rev Med Devices. 2017. PMID: 28276752 Review.
Cited by
-
Recent advances in cardiac SPECT instrumentation and system design.Curr Cardiol Rep. 2013 Aug;15(8):387. doi: 10.1007/s11886-013-0387-x. Curr Cardiol Rep. 2013. PMID: 23832650 Review.
-
Diagnosing coronary artery disease with hybrid PET/CT: it takes two to tango.J Nucl Cardiol. 2013 Oct;20(5):874-90. doi: 10.1007/s12350-013-9753-8. J Nucl Cardiol. 2013. PMID: 23842709 Review.
-
Feasibility of ultra low-dose thallium stress-redistribution protocol including prone imaging in obese patients using CZT camera.Int J Cardiovasc Imaging. 2016 Sep;32(9):1463-1469. doi: 10.1007/s10554-016-0919-8. Epub 2016 Jun 2. Int J Cardiovasc Imaging. 2016. PMID: 27250348
-
Downstream resource utilization following hybrid cardiac imaging with an integrated cadmium-zinc-telluride/64-slice CT device.Eur J Nucl Med Mol Imaging. 2012 Mar;39(3):430-6. doi: 10.1007/s00259-011-1999-2. Epub 2011 Dec 6. Eur J Nucl Med Mol Imaging. 2012. PMID: 22143224
-
Nuclear cardiology: 2014 innovations and developments.Methodist Debakey Cardiovasc J. 2014 Jul-Sep;10(3):163-71. doi: 10.14797/mdcj-10-3-163. Methodist Debakey Cardiovasc J. 2014. PMID: 25574344 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical