Vertebroplasty: only small cement volumes are required to normalize stress distributions on the vertebral bodies
- PMID: 20010394
- DOI: 10.1097/BRS.0b013e3181b4ea1e
Vertebroplasty: only small cement volumes are required to normalize stress distributions on the vertebral bodies
Abstract
STUDY DESIGN.: Biomechanical study of vertebroplasty in cadaver motion segments. OBJECTIVES.: To determine how the volume of injected cement influences: (a) stress distributions on fractured and adjacent vertebral bodies, (b) load-sharing between the vertebral bodies and neural arch, and (c) cement leakage. SUMMARY OF BACKGROUND DATA.: Vertebroplasty is increasingly used to treat vertebral fractures, but there are problems concerning adjacent level fracture and cement leakage, both of which may depend on the volume of injected cement. METHODS.: Nineteen thoracolumbar motion segments from 13 cadavers (42-91 years) were loaded to induce fracture. Fractured vertebrae received 2 sequential injections (VP1 and VP2) of 3.5 cm of polymethylmethacrylate cement. Before and after each intervention, motion segment stiffness was measured in compression and in bending, and "stress profilometry" was used to quantify the distribution of compressive stress in the intervertebral disc (which presses equally on fractured and adjacent vertebrae). Stress profiles were obtained by pulling a pressure transducer through the disc while the motion segment was compressed in flexed and extended postures. Stress profiles yielded the intradiscal pressure (IDP), the magnitude of stress peaks in the anterior and posterior (SPP) anulus, and the percentage of the applied compressive force resisted by the neural arch (FN). Cement leakage and vertebral body volume were quantified using water-immersion, and the percentage cement fill was estimated. RESULTS.: Bending and compressive stiffness fell by 37% and 50% respectively following fracture, and were restored only after VP2. Depending on posture, IDP fell by 59-85% after fracture whereas SPP increased by 107- 362%. VP1 restored IDP and SPP to prefracture values, and VP2 produced no further changes. Fracture increased FN from 11% to 39% in flexion, and from 33% to 59% in extension. FN was restored towards prefracture values only after VP2. Cement leakage increased after VP2 and was negatively correlated to vertebral body volume. Following VP2, increases in IDP and compressive stiffness were proportional to percentage fill. CONCLUSION.: About 3.5 cm of PMMA largely restored normal stress distributions to fractured and adjacent vertebral bodies, but 7 cm were required to restore motion segment stiffness and load-sharing between the vertebral bodies and neural arch. Cement leakage, IDP and compressive stiffness all increased with percentage fill.
Similar articles
-
Is kyphoplasty better than vertebroplasty in restoring normal mechanical function to an injured spine?Bone. 2010 Apr;46(4):1050-7. doi: 10.1016/j.bone.2009.11.036. Epub 2009 Dec 11. Bone. 2010. PMID: 20004264
-
Mechanical efficacy of vertebroplasty: influence of cement type, BMD, fracture severity, and disc degeneration.Bone. 2007 Apr;40(4):1110-9. doi: 10.1016/j.bone.2006.11.021. Epub 2007 Jan 16. Bone. 2007. PMID: 17229596
-
Effect of bone density on vertebral strength and stiffness after percutaneous vertebroplasty.Spine (Phila Pa 1976). 2007 Aug 15;32(18):E505-11. doi: 10.1097/BRS.0b013e318133fc73. Spine (Phila Pa 1976). 2007. PMID: 17700430
-
[Latest advances of the cause and preventive strategies of adjacent vertebral body fracture after percutaneous vertebroplasty].Zhongguo Gu Shang. 2010 Oct;23(10):792-5. Zhongguo Gu Shang. 2010. PMID: 21137300 Review. Chinese.
-
Combined extraforaminal and intradiscal cement leakage following percutaneous vertebroplasty.Spine (Phila Pa 1976). 2007 May 20;32(12):E358-62. doi: 10.1097/BRS.0b013e31805c0b25. Spine (Phila Pa 1976). 2007. PMID: 17515810 Review.
Cited by
-
Cement volume is the most important modifiable predictor for pain relief in BKP: results from SWISSspine, a nationwide registry.Eur Spine J. 2013 Oct;22(10):2241-8. doi: 10.1007/s00586-013-2869-3. Epub 2013 Jun 18. Eur Spine J. 2013. PMID: 23775292 Free PMC article.
-
Clinical Effects and Radiological Results of Vertebroplasty: Over a 2-year Follow-Up Period.Korean J Spine. 2012 Dec;9(4):334-9. doi: 10.14245/kjs.2012.9.4.334. Epub 2012 Dec 31. Korean J Spine. 2012. PMID: 25983842 Free PMC article.
-
Repeat vertebroplasty for the subsequent refracture of procedured vertebra.Korean J Pain. 2013 Jan;26(1):94-7. doi: 10.3344/kjp.2013.26.1.94. Epub 2013 Jan 4. Korean J Pain. 2013. PMID: 23342217 Free PMC article.
-
Volume matters: a review of procedural details of two randomised controlled vertebroplasty trials of 2009.Eur Spine J. 2010 Nov;19(11):1837-40. doi: 10.1007/s00586-010-1525-4. Epub 2010 Aug 5. Eur Spine J. 2010. PMID: 20686793 Free PMC article. Review.
-
Comparative clinical results of vertebroplasty using jamshidi® needle and bone void filler for acute vertebral compression fractures.Korean J Spine. 2012 Sep;9(3):239-43. doi: 10.14245/kjs.2012.9.3.239. Epub 2012 Sep 30. Korean J Spine. 2012. PMID: 25983822 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous