Overestimation of low-dose radiation in intensity-modulated radiotherapy with sliding-window technique
- PMID: 20013092
- DOI: 10.1007/s00066-009-2028-2
Overestimation of low-dose radiation in intensity-modulated radiotherapy with sliding-window technique
Abstract
Purpose: To analyze different control-system limitations on the measured dose distributions in low-dose regions of simplified intensity fields with an electronic portal imaging device to ascertain the optimal settings for the control-system limitations in the planning system.
Material and methods: The authors created one field with an "optimal fluence" of intensity 1.0 (full dose) and one field with intensity 0.0 (no dose) in the central part of the field. The influence of different dose rates (DRs) and maximum leaf speeds (LS) on the calculated and measured dose and dose profiles were analyzed.
Results: Good agreement between calculated and measured dose in the case of a field of intensity 1.0 was found. For the field with intensity 0.0, the measured dose was 20-60% lower than the dose calculated by the "actual fluence". The results were found dependent on the DR and LS.
Conclusion: The overestimation in regions of optimal intensity 0.0 by the planning system cannot be resolved by the user. Taking the measured dose in the region of desired intensity 1.0 and other technical limitations (like beam hold interrupts or spikes in the cross and longitudinal profiles) into consideration, the application of an LS of 2.5 cm/s and a DR of 500 MU/min is recommended in order to minimize radiation dose applied to organs at risk, which are located in regions of low intensity, like, for example, the spinal cord.
Similar articles
-
Impact of different leaf velocities and dose rates on the number of monitor units and the dose-volume-histograms using intensity modulated radiotherapy with sliding-window technique.Radiat Oncol. 2008 Sep 23;3:31. doi: 10.1186/1748-717X-3-31. Radiat Oncol. 2008. PMID: 18811954 Free PMC article.
-
A topographic leaf-sequencing algorithm for delivering intensity modulated radiation therapy.Med Phys. 2006 Aug;33(8):2751-6. doi: 10.1118/1.2216876. Med Phys. 2006. PMID: 16964850
-
An empirical model of electronic portal imager response implemented within a commercial treatment planning system for verification of intensity-modulated radiation therapy fields.J Appl Clin Med Phys. 2008 Nov 11;9(4):135-150. doi: 10.1120/jacmp.v9i4.2807. J Appl Clin Med Phys. 2008. PMID: 19020485 Free PMC article.
-
SWIMRT: a graphical user interface using sliding window algorithm to construct a fluence map machine file.J Appl Clin Med Phys. 2006 May 25;7(2):69-85. doi: 10.1120/jacmp.v7i2.2231. J Appl Clin Med Phys. 2006. PMID: 17533330 Free PMC article.
-
Monte Carlo implementation, validation, and characterization of a 120 leaf MLC.Med Phys. 2011 Oct;38(10):5311-20. doi: 10.1118/1.3626485. Med Phys. 2011. PMID: 21992349
Cited by
-
Helical tomotherapy in cervical cancer patients: simultaneous integrated boost concept: technique and acute toxicity.Strahlenther Onkol. 2010 Oct;186(10):572-9. doi: 10.1007/s00066-010-2121-6. Epub 2010 Sep 30. Strahlenther Onkol. 2010. PMID: 20936458
-
Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules.Strahlenther Onkol. 2011 May;187(5):316-21. doi: 10.1007/s00066-011-2176-z. Epub 2011 Apr 26. Strahlenther Onkol. 2011. PMID: 21544527
-
A planning comparison of dynamic IMRT for different collimator leaf thicknesses with helical tomotherapy and RapidArc for prostate and head and neck tumors.Strahlenther Onkol. 2010 Sep;186(9):502-10. doi: 10.1007/s00066-010-2124-3. Epub 2010 Aug 30. Strahlenther Onkol. 2010. PMID: 20803184
References
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials