Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T

Affiliations
Comparative Study

Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T

Meike Becker et al. Eur Radiol. 2010 Jun.

Abstract

Objective: As high-field cardiac MRI (CMR) becomes more widespread the propensity of ECG to interference from electromagnetic fields (EMF) and to magneto-hydrodynamic (MHD) effects increases and with it the motivation for a CMR triggering alternative. This study explores the suitability of acoustic cardiac triggering (ACT) for left ventricular (LV) function assessment in healthy subjects (n = 14).

Methods: Quantitative analysis of 2D CINE steady-state free precession (SSFP) images was conducted to compare ACT's performance with vector ECG (VCG). Endocardial border sharpness (EBS) was examined paralleled by quantitative LV function assessment.

Results: Unlike VCG, ACT provided signal traces free of interference from EMF or MHD effects. In the case of correct R-wave recognition, VCG-triggered 2D CINE SSFP was immune to cardiac motion effects-even at 3.0 T. However, VCG-triggered 2D SSFP CINE imaging was prone to cardiac motion and EBS degradation if R-wave misregistration occurred. ACT-triggered acquisitions yielded LV parameters (end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF) and left ventricular mass (LVM)) comparable with those derived from VCG-triggered acquisitions (1.5 T: ESV(VCG) = (56 +/- 17) ml, EDV(VCG) = (151 +/- 32) ml, LVM(VCG) = (97 +/- 27) g, SV(VCG) = (94 +/- 19) ml, EF(VCG) = (63 +/- 5)% cf. ESV(ACT) = (56 +/- 18) ml, EDV(ACT) = (147 +/- 36) ml, LVM(ACT) = (102 +/- 29) g, SV(ACT) = (91 +/- 22) ml, EF(ACT) = (62 +/- 6)%; 3.0 T: ESV(VCG) = (55 +/- 21) ml, EDV(VCG) = (151 +/- 32) ml, LVM(VCG) = (101 +/- 27) g, SV(VCG) = (96 +/- 15) ml, EF(VCG) = (65 +/- 7)% cf. ESV(ACT) = (54 +/- 20) ml, EDV(ACT) = (146 +/- 35) ml, LVM(ACT) = (101 +/- 30) g, SV(ACT) = (92 +/- 17) ml, EF(ACT) = (64 +/- 6)%).

Conclusions: ACT's intrinsic insensitivity to interference from electromagnetic fields renders it suitable for clinical CMR.

PubMed Disclaimer

Similar articles

Cited by

References

    1. AJR Am J Roentgenol. 2008 Oct;191(4):1208-15 - PubMed
    1. J Magn Reson Imaging. 2008 Sep;28(3):767-72 - PubMed
    1. J Magn Reson Imaging. 2007 Nov;26(5):1296-302 - PubMed
    1. J Magn Reson Imaging. 1996 Jan-Feb;6(1):271-2 - PubMed
    1. Magn Reson Med. 2002 Sep;48(3):425-9 - PubMed

LinkOut - more resources