Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T
- PMID: 20013275
- DOI: 10.1007/s00330-009-1676-z
Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T
Abstract
Objective: As high-field cardiac MRI (CMR) becomes more widespread the propensity of ECG to interference from electromagnetic fields (EMF) and to magneto-hydrodynamic (MHD) effects increases and with it the motivation for a CMR triggering alternative. This study explores the suitability of acoustic cardiac triggering (ACT) for left ventricular (LV) function assessment in healthy subjects (n = 14).
Methods: Quantitative analysis of 2D CINE steady-state free precession (SSFP) images was conducted to compare ACT's performance with vector ECG (VCG). Endocardial border sharpness (EBS) was examined paralleled by quantitative LV function assessment.
Results: Unlike VCG, ACT provided signal traces free of interference from EMF or MHD effects. In the case of correct R-wave recognition, VCG-triggered 2D CINE SSFP was immune to cardiac motion effects-even at 3.0 T. However, VCG-triggered 2D SSFP CINE imaging was prone to cardiac motion and EBS degradation if R-wave misregistration occurred. ACT-triggered acquisitions yielded LV parameters (end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF) and left ventricular mass (LVM)) comparable with those derived from VCG-triggered acquisitions (1.5 T: ESV(VCG) = (56 +/- 17) ml, EDV(VCG) = (151 +/- 32) ml, LVM(VCG) = (97 +/- 27) g, SV(VCG) = (94 +/- 19) ml, EF(VCG) = (63 +/- 5)% cf. ESV(ACT) = (56 +/- 18) ml, EDV(ACT) = (147 +/- 36) ml, LVM(ACT) = (102 +/- 29) g, SV(ACT) = (91 +/- 22) ml, EF(ACT) = (62 +/- 6)%; 3.0 T: ESV(VCG) = (55 +/- 21) ml, EDV(VCG) = (151 +/- 32) ml, LVM(VCG) = (101 +/- 27) g, SV(VCG) = (96 +/- 15) ml, EF(VCG) = (65 +/- 7)% cf. ESV(ACT) = (54 +/- 20) ml, EDV(ACT) = (146 +/- 35) ml, LVM(ACT) = (101 +/- 30) g, SV(ACT) = (92 +/- 17) ml, EF(ACT) = (64 +/- 6)%).
Conclusions: ACT's intrinsic insensitivity to interference from electromagnetic fields renders it suitable for clinical CMR.
Similar articles
-
Cardiac MRI: evaluation of phonocardiogram-gated cine imaging for the assessment of global und regional left ventricular function in clinical routine.Eur Radiol. 2012 Mar;22(3):559-68. doi: 10.1007/s00330-011-2279-z. Epub 2011 Sep 24. Eur Radiol. 2012. PMID: 21947482
-
Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla.J Cardiovasc Magn Reson. 2010 Nov 16;12(1):67. doi: 10.1186/1532-429X-12-67. J Cardiovasc Magn Reson. 2010. PMID: 21080933 Free PMC article.
-
Feasibility of cardiac gating free of interference with electro-magnetic fields at 1.5 Tesla, 3.0 Tesla and 7.0 Tesla using an MR-stethoscope.Invest Radiol. 2009 Sep;44(9):539-47. doi: 10.1097/RLI.0b013e3181b4c15e. Invest Radiol. 2009. PMID: 19652614
-
Free breathing 2D multi-slice real-time gradient-echo cardiovascular magnetic resonance imaging: impact on left ventricular function measurements compared with standard multi-breath hold 2D steady-state free precession imaging.Acta Cardiol. 2011 Aug;66(4):489-97. doi: 10.1080/ac.66.4.2126598. Acta Cardiol. 2011. PMID: 21894806
-
Detection of cardiovascular disease in elite athletes using cardiac magnetic resonance imaging.Rofo. 2013 Dec;185(12):1167-74. doi: 10.1055/s-0033-1350130. Epub 2013 Jul 29. Rofo. 2013. PMID: 23897528 Review.
Cited by
-
Accelerated fast spin-echo magnetic resonance imaging of the heart using a self-calibrated split-echo approach.PLoS One. 2014 Apr 11;9(4):e94654. doi: 10.1371/journal.pone.0094654. eCollection 2014. PLoS One. 2014. PMID: 24728341 Free PMC article.
-
Self-gated fat-suppressed cardiac cine MRI.Magn Reson Med. 2015 May;73(5):1764-74. doi: 10.1002/mrm.25291. Epub 2014 May 7. Magn Reson Med. 2015. PMID: 24806049 Free PMC article.
-
Quantification of Right and Left Ventricular Function in Cardiac MR Imaging: Comparison of Semiautomatic and Manual Segmentation Algorithms.Diagnostics (Basel). 2013 Apr 3;3(2):271-82. doi: 10.3390/diagnostics3020271. Diagnostics (Basel). 2013. PMID: 26835680 Free PMC article.
-
Cardiac magnetic resonance imaging and its electrocardiographs (ECG): tips and tricks.Int J Cardiovasc Imaging. 2012 Aug;28(6):1465-75. doi: 10.1007/s10554-011-9957-4. Epub 2011 Oct 28. Int J Cardiovasc Imaging. 2012. PMID: 22033762 Free PMC article. Review.
-
Doppler Ultrasound Triggering for Cardiovascular MRI at 3T in a Healthy Volunteer Study.Magn Reson Med Sci. 2017 Apr 10;16(2):98-108. doi: 10.2463/mrms.mp.2015-0104. Epub 2016 Mar 21. Magn Reson Med Sci. 2017. PMID: 27001390 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials