Comparison of plethysmographic and helium dilution lung volumes: which is best for COPD?
- PMID: 20022972
- PMCID: PMC2862399
- DOI: 10.1378/chest.09-1504
Comparison of plethysmographic and helium dilution lung volumes: which is best for COPD?
Abstract
Background: Theoretical considerations and limited scientific evidence suggest that whole-body plethysmography overestimates lung volume in patients with severe airflow obstruction. We sought to compare plethysmography (Pleth)-, helium dilution (He)- and CT scan-derived lung volume measurements in a sample containing many patients with severe airflow obstruction.
Methods: We measured total lung capacity (TLC) in 132 patients at three hospitals, with monitored application of recommended techniques for Pleth and He measurements of lung volume and by thoracic CT scans obtained during breath hold at full inspiration.
Results: Average TLC among 132 subjects was 6.18 L (+/- 1.69 L) by Pleth-derived TLC, 5.55 L (+/- 1.39 L) by He-derived TLC, and 5.31 L (+/- 1.47) by CT scan-derived TLC. Pleth-derived TLC was significantly greater than either He-derived TLC or CT scan-derived TLC (P < or = .001), whereas there was no significant difference between He-derived and CT scan-derived values. When examined separately, there were significant within-subject differences in TLC by measurement technique among subjects with airflow obstruction, but not among those without airflow obstruction. Plethysmographic overestimation of TLC was greatest among subjects with FEV(1) < 30% of predicted.
Conclusions: In the setting of airflow obstruction, Pleth systematically overestimates lung volume relative to He or thoracic imaging despite adherence to current recommendations for proper measurement technique.
Figures
Comment in
-
Lung volumes in COPD: not only the total lung capacity.Chest. 2010 Jul;138(1):233; author reply 233-4. doi: 10.1378/chest.10-0045. Chest. 2010. PMID: 20605829 No abstract available.
-
Comparison among different methods on measurement of total lung capacity in COPD.Chest. 2010 Aug;138(2):458-9; author reply 459-60. doi: 10.1378/chest.10-0369. Chest. 2010. PMID: 20682542 No abstract available.
-
How to measure lung volume?Chest. 2010 Nov;138(5):1280-1; author reply 1281-2. doi: 10.1378/chest.10-1369. Chest. 2010. PMID: 21051413 No abstract available.
References
-
- Schaanning CG, Gulsvik A. Accuracy and precision of helium dilution technique and body plethysmography in measuring lung volumes. Scand J Clin Lab Invest. 1973;32(3):271–277. - PubMed
-
- Quanjer PH, Tammeling GJ, Cotes JE, Pedersen OF, Peslin R, Yernault JC, Official Statement of the European Respiratory Society Lung volumes and forced ventilatory flows. Report Working Party Standardization of Lung Function Tests, European Community for Steel and Coal. Official statement of the European Respiratory Society. Eur Respir J Suppl. 1993;16:5–40. - PubMed
-
- Wanger J, Clausen JL, Coates A, et al. Standardisation of the measurement of lung volumes. Eur Respir J. 2005;26(3):511–522. - PubMed
-
- Stănescu DC, Rodenstein DO. Which method for measuring lung volume in airway obstruction [abstract] Presented at American Thoracic Society Annual Meeting; May, 1982.
-
- Rodenstein DO, Francis C, Stănescu DC. Airway closure in humans does not result in overestimation of plethysmographic lung volume. J Appl Physiol. 1983;55(6):1784–1789. - PubMed
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical