Abstract
Background: Parathyroid hormone (PTH) assays able to distinguish between full-length PTH (PTH1-84) and N-terminally truncated PTH (PTH7-84) are of increasing significance in the accurate diagnosis of endocrine and osteological diseases. We describe the discovery of new N-terminal and C-terminal PTH variants and the development of selected reaction monitoring (SRM)-based immunoassays specifically designed for the detection of full-length PTH [amino acid (aa)1-84] and 2 N-terminal variants, aa7-84 and aa34-84.
Methods: Preparation of mass spectrometric immunoassay pipettor tips and MALDI-TOF mass spectrometric analysis were carried out as previously described. We used novel software to develop SRM assays on a triple-quadrupole mass spectrometer. Heavy isotope-labeled versions of target peptides were used as internal standards.
Results: Top-down analysis of samples from healthy individuals and renal failure patients revealed numerous PTH variants, including previously unidentified aa28-84, aa48-84, aa34-77, aa37-77, and aa38-77. Quantitative SRM assays were developed for PTH1-84, PTH7-84, and variant aa34-84. Peptides exhibited linear responses (R(2) = 0.90-0.99) relative to recombinant human PTH concentration limits of detection for intact PTH of 8 ng/L and limits of quantification of 16-31 ng/L depending on the peptide. Standard error of analysis for all triplicate measurements was 3%-12% for all peptides, with <5% chromatographic drift between replicates. The CVs of integrated areas under the curve for 54 separate measurements of heavy peptides were 5%-9%.
Conclusions: Mass spectrometric immunoassays identified new clinical variants of PTH and provided a quantitative assay for these and previously identified forms of PTH.
Similar articles
-
Chemical Characterization and Quantification of Circulating Intact PTH and PTH Fragments by High-Resolution Mass Spectrometry in Chronic Renal Failure.Clin Chem. 2021 Jun 1;67(6):843-853. doi: 10.1093/clinchem/hvab013. Clin Chem. 2021. PMID: 33693557 Free PMC article.
-
Quantification and clinical performance of serum parathyroid hormone 1-84 via immunocapture coupled to LC-MS/MS in chronic renal failure.J Pharm Biomed Anal. 2025 Apr 15;256:116678. doi: 10.1016/j.jpba.2025.116678. Epub 2025 Jan 14. J Pharm Biomed Anal. 2025. PMID: 39842077
-
Immunoassays for the detection of parathyroid hormone.J Bone Miner Res. 2002 Nov;17 Suppl 2:N81-6. J Bone Miner Res. 2002. PMID: 12412782 Review.
-
Quantification of serum 1-84 parathyroid hormone in patients with hyperparathyroidism by immunocapture in situ digestion liquid chromatography-tandem mass spectrometry.Clin Chem. 2010 Feb;56(2):306-13. doi: 10.1373/clinchem.2009.134643. Epub 2009 Dec 10. Clin Chem. 2010. PMID: 20007860
-
LC-MS candidate reference methods for the harmonisation of parathyroid hormone (PTH) measurement: a review of recent developments and future considerations.Clin Chem Lab Med. 2014 Sep;52(9):1251-63. doi: 10.1515/cclm-2014-0150. Clin Chem Lab Med. 2014. PMID: 24762644 Review.
Cited by
-
Parathyroid Hormone Measurement in Chronic Kidney Disease: From Basics to Clinical Implications.Int J Nephrol. 2019 Sep 17;2019:5496710. doi: 10.1155/2019/5496710. eCollection 2019. Int J Nephrol. 2019. PMID: 31637056 Free PMC article. Review.
-
Mass spectrometric immunoassay revisited.J Am Soc Mass Spectrom. 2011 Jun;22(6):960-8. doi: 10.1007/s13361-011-0094-z. Epub 2011 Mar 8. J Am Soc Mass Spectrom. 2011. PMID: 21953037 Free PMC article. Review.
-
Development of a Chip/Chip/SRM platform using digital chip isoelectric focusing and LC-Chip mass spectrometry for enrichment and quantitation of low abundance protein biomarkers in human plasma.J Proteome Res. 2012 Feb 3;11(2):808-17. doi: 10.1021/pr2006704. Epub 2011 Dec 27. J Proteome Res. 2012. PMID: 22098410 Free PMC article.
-
Replacing immunoassays with tryptic digestion-peptide immunoaffinity enrichment and LC-MS/MS.Bioanalysis. 2012 Feb;4(3):281-90. doi: 10.4155/bio.11.319. Bioanalysis. 2012. PMID: 22303832 Free PMC article. Review.
-
Mass spectrometric immunoassays for discovery, screening and quantification of clinically relevant proteoforms.Bioanalysis. 2016 Aug;8(15):1623-1633. doi: 10.4155/bio-2016-0060. Epub 2016 Jul 11. Bioanalysis. 2016. PMID: 27396364 Free PMC article. Review.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials
Miscellaneous